Current results
If you want to export current results of analysis please follow the instruction below.
1. Select wells that you want to get results for.
2. Choose if you want to analyze per Target or per Channel by clicking proper tab, and select
targets or channels you are interested in.
3. Click on the Show results button.
4. In the List tab, above the table, on the right-hand side you can see the Export to CSV... option.
5. Click on the Export to CSV option and select the Current results option.
6. Remember that data in CSV reflects data from List View are structured as follows:
Well – identifier of well
Sample/NTC/Control – name of Sample/NTC/Control
Reaction mix – name of reaction mix
Channel – channel name or Target – target name
IC – internal control if selected
Control Type – control type
Concentration (copies/ µl) – calculated conentration
CI (%) – confidence interval parameter
Partitions (valid) – number of valid partitions
Partitions (positive) – number of positive partitions
Partitions (negative) – number of negative partitions
Threshold – value of threshold
Multiple occupancy
If you want to export a number of multiple occupancies for multiplex reaction, please follow the
instructions below.
1. Select wells that you want to get results for.
2. Click on the Channels tab, and select channels you are interested in.
3. Click on the Show results button.
4. In the List tab, above the table, on the right-hand side, you can see the Export to CSV... option.
5. Click on the Export to CSV option and select the Multiple occupancy option.
6. Remember that data in CSV are structured as follows:
Well – identifier of well
Hyperwell – identifier of hyperwell if wells were grouped
QIAcuity User Manual 06/2022