The QIAcuity instrument(s) and QIAcuity Software Suite are installed to a network, allowing multiple clients to access the QIAcuity
instrument(s) via a single QIAcuity Suite server. For connection to the Suite server, use the IP address of the Suite server.
Configuring an Ethernet connection between the QIAcuity instrument and the QIAcuity
Software Suite
To establish a connection, the instrument and the notebook must be connected to the LAN (local
area network). For 10 users accessing the system in parallel, the requirement is a minimum data
connection speed of 10Mbit/s for a setup via network
Note: Please make sure that the QIAcuity instrument is connected to the LAN. Any other
configurations are not supported by QIAGEN.
Note: Only users with Administrator role can modify the network configuration. We recommend to
consult your network administrator when configuring the network. For communication with QIAcuity
Software Suite, the following ports are used:
The inbound TCP port 44321
The inbound TCP port 8080
The inbound TCP port 8687
The inbound UDP port 9595
Pinging the network is also supported.
QIAcuity User Manual 06/2022
52