2. Edit Plate Data — User can check and edit parameters of owned Plate (dPCR parameters,
plate layout (samples, reaction mixes (reagents), controls), and mark it as primed.
Note: 'Read owned Plate' required to have fully available permission.
3. Edit Analysis Data — User can change the threshold and use lasso selection on the Analysis
page of the owned Plates to verify the accuracy of the results.
4. Read Plate — User can search the Plate, analyze the Plate, see all created Plates, check
details about owned Plate (dPCR parameters, plate layout (samples, reaction mixes, controls)),
and export results to CSV.
5. Delete Plate — User can delete owned Plates.
Note: 'Read owned Plate' required to have fully available permission.
Other permissions
1. Import Plate — User can import the Plate as a ZIP file.
Note: 'Read owned Plate' or 'Read all Plate' required to have fully available permission.
Import plate
The import plate button is visible in Plates overview.
2. Export Plate — User can export the Plate as a ZIP file.
Note: 'Read owned Plate' or 'Read all Plate' required to have fully available permission.
Export plate
The export plate button is visible after clicking the three dots (...) on the plate.
The export plate button is visible in plate details view.
3. Unlock Plate — User can unlock locked Plate.
Note: 'Read owned Plate' or 'Read all Plate' required to have fully available permission.
Unlock plate
When a Plate is in locked status, then the unlock plate button is visible in plate details.
Note: Only tap the unlock plate button in case the run was stopped by an instrument
error and the instrument is not unlocking the plate, although it was taken out from the
QIAcuity instrument. If the Unlock plate button is tapped during a run, the run will be
stopped and the data will get lost.
4. Set Plate Ownership — User can set owners of the plate.
Note: 'Read owned Plate' or 'Read all Plate' and 'Edit owned Plate' or 'Edit all Plate' required
to have fully available permission.
Set plate ownership
QIAcuity User Manual 06/2022