Qiagen QIAcuity 911000 User Manual page 281

Hide thumbs Also See for QIAcuity 911000:
Table of Contents

Advertisement

g)
High copy number in NTC
h)
Lower RFU of negative partitions in
NTC/samples with low number of
positive partitions
i)
The confidence interval is wide
j)
Vertical stripes in the images
k)
Double positive or double negative
signals
l)
Fetch error while accessing the Users
list in User Management
Software
a)
The QIAcuity Software Suite does not
start
b)
Disc space is critical in the QIAcuity
Software Suite
c)
User forgot password
d)
Communication error between
QIAcuity instrument and software
e)
Instrument software or Software Suite
is unresponsive
f)
Startup of instruments displays an
error
g)
Error 205 or Error 32
h)
Error 490
QIAcuity User Manual 06/2022
Check the images or signal map for dust or other particles. In
case, wipe the plate with a lint-free tissue (optionally, use ethanol)
and re-image the plate.
The signal intensity might be lower in images with high number of
negative partitions. There's no influence on the result analysis, as
the signal to noise is not affected.
The number of valid partitions is low. Check the images for
blacked-out areas, that can occur, e.g., due to bad filling or
areas of low amplification (see section Image corrective
measures)
Please re-image the plate for proper image analysis
The double positive or double negative signals could have
different root causes. One of the reasons for observing double
signal bands could be suboptimal assay designs, such as cross-
hybridization of probes to unspecific targets or secondary off-
target amplification products. Besides assays-related causes,
improper crosstalk compensation could also be the root cause.
An insufficient compensation or overcompensation of crosstalk
from neighboring channels could also result in extra signal
bands. To determine the main root cause, re-image the plate with
30% less exposure times for affected channel. If double bands
disappear or get much closer to each other after re-imaging, they
are most likely to be caused by improper crosstalk compensation
rather than assay-related issues.
If such error occurs, please contact QIAGEN Technical Service in
order to solve the issue.
Check that the software is installed on the laptop. Check the
operating system. The QIAcuity Software Suite can only be
operated and Windows 10.
Delete plates from the plates overview.
Administrator to log-in, change the password for the user. If
administrator forgot the password, contact QIAGEN Technical
Service.
This error occurs when the data received from the instrument does
not conform to the expected pattern.
Further investigations are required by a QIAGEN Field Service
Specialist to diagnose the problem with the instrument.
Please contact your distributor or QIAGEN Technical Service.
Re-start the QIAcuity instrument or the notebook where the
QIAcuity Software Suite is installed
The required plate recovery task cannot be performed because
there are no plate slots available in the tray. Remove all loaded
plates before you proceed. Press Restart to start recovery.
The error can occur in different situations:
A) Please make sure that the selected Plate type corresponds with
the entered barcode, if manually entered. If not matching, it will
lead to an error on the instrument (error 205).
B) Please make sure that after the first successful suite connection,
the instrument is restarted to allow automatic synchronization of
labware files.
The error can occur after a plate was processed and a failure in
image transfer to the Suite was detected. The Suite rejected a
data package due to improper format. Please see if all images
are available in the Suite. If you find images missing, add an
additional imaging step to recover the data.