Synchronizing labware with the QIAcuity Software Suite
The QIAcuity Control Software and the QIAcuity Software Suite use labware files that include
information about the nanoplate formats available. A labware file defines dimensional specification
and default processing parameters for each Nanoplate. Information about the labware used by the
QIAcuity can be downloaded from the Software Suite. The labware data is synchronized
automatically when the instrument is turned on and the connection to the Suite is established. You
can also synchronize it manually by clicking Tools and selecting Labware Sync. The labware is then
downloaded and updated.
Note: Only user with appropriate rights are allowed to synchronize labware.
Note: Please make sure that after the first successful suite connection, the instrument is restarted to
allow automatic synchronization of labware files. In case error 205 or 32 is shown, manually
synchronize labware within the tools menu and restart the instrument.
Note: The connection between the QIAcuity and the Software Suite must be configured before
starting this procedure. For more information on how to set up the connection, refer to section
Establishing a connection between the QIAcuity instrument and the QIAcuity Software Suite.
To download labware information from the Software Suite, follow the steps below.
1. On the toolbar, tap Tools.
2. Tap Servicing.
3. Tap Labware sync. The Labware synchronization dialog box appears.