Qiagen QIAcuity 911000 User Manual page 283

Hide thumbs Also See for QIAcuity 911000:
Table of Contents

Advertisement

System status environment on the QIAcuity Eight after an error occurs.
4. To clear an error, tap Clear error.
5. If the error that occurred affects the tray(s), tap More details. To clear a tray-related error, tap
Clear error in the dialog. The dialog box contains five items that can be cleared for each tray,
such as motor and slot numbers (based on Insrument Version).
Note: In QIAcuity Eight, the More details button is located in the Upper Tray and Lower Tray
panes. In the QIAcuity Four and QIAcuity One, the More details button is located in the Tray
pane.
6. Restart the instrument. The instrument initializes and all modules are returned to their home
positions.
Note: If the affected module is not working after you cleared the error and restarted the
instrument, contact QIAGEN Technical Services.
QIAcuity User Manual 06/2022