The following actions can be performed when your run is finished:
To reimage the plate with different settings, click +, then select Imaging.
To rerun the plate with additional cycles, click +, then select Cycling + Imaging
Important: After a run is finished, remove the plate from the instrument before adding a new
imaging or cycling + imaging step.
Note: For QIAcuity Software Suite v2.0 or higher dust detected in one channel is used for finding
dust in other channels as well. If required, to improve dust detection always image all channels
even not used by assay(s).
Defining reaction mixes
You can create a new reaction mix or use an existing reaction mix for your experiment.
The General View tab displays the name of all reaction mixes stored in your database. Click the
three-dotted icon of the reaction mix if you want to edit or delete the reaction mix.
Important: Modifying an existing reaction mix changes the settings of the reaction mix in all
experiments where it is used.