Qiagen QIAcuity 911000 User Manual page 21

Hide thumbs Also See for QIAcuity 911000:
Table of Contents

Advertisement

Power switch
The main power switch is located at the back of the QIAcuity. To power ON the QIAcuity, turn the
power switch to I and press the blue soft-switch button at the front of the instrument. The startup
screen appears, and the instrument automatically performs initialization tests.
To conserve energy, the QIAcuity can be powered OFF when not in use. To power OFF the
QIAcuity, press the blue front soft-switch.
Important: After powering OFF the QIAcuity, wait for a few seconds before switching ON the
instrument again. The system might fail to start if you do not allow the QIAcuity to rest for a few
seconds before powering ON.
RJ-45 Ethernet port
The RJ-45 Ethernet port is located at the back of the instrument beside the power cord socket. It is
only used to connect the QIAcuity to local area network via cable.
USB ports
The QIAcuity has two USB ports that are located at the front of the instrument in the upper left
corner. A third slot for accessories is available behind the touchscreen in the upper right corner. To
access this slot, extend the touchscreen as far as possible.
The USB ports allow a connection of the QIAcuity to a USB stick. Data files, such as support
package, can be transferred via the USB port from the QIAcuity to the USB stick. The USB ports can
also be used to plug in an external bar code reader or a mouse/keyboard.
Important: We recommend using QIAGEN USB sticks only to ensure full compatibility.
Important: Do not remove the USB stick while downloading or transferring data or software to or
from the instrument.
Power cord socket
The power cord socket is located at the rear right of the QIAcuity and allows connection of the
QIAcuity to a power outlet via the supplied power cord.
QIAcuity User Manual 06/2022
21

Advertisement

Table of Contents
loading

This manual is also suitable for:

Qiacuity 911020Qiacuity 911040Qiacuity 911050

Table of Contents