Peripheral Devices

12.6.4 Battery Mode

12.6.4.1 Battery Life Cycle

12-12
Index
Battery
status
03
Capacity
Low
04
Capacity
shutdown
05
Shutdown
10% < x < 15%
06
Shutdown
About 10% < x < 15%
System
When in Battery Power Mode, the system maintains the current patient and can scan or
perform other operation normally.
Battery life cycle can estimate using cycle count. If cycle count reaches up to 250 cycles,
Battery will be expired.
Capacity ± 1,770mAh @ after 250 cycles (60% of the standard capacity @RT)
- Charge: 1.5A 4.20V, CCCV 150mA cut-off @ RT.
-Discharge: 15A, 2.5V cut-off @ RT
Assumption : 1cycle/day
For example, user uses function and discharge 1 cycle per day, battery lifecycle is 250
days (in workding day).
For example, user uses battery function and discharge 0.5 cycle per day, battery
lifecycle is 500 days (in workding day).
Definition of Cycle
Battery is monitoring total discharge-capacity for Cycle-count.
1 cycle = total usage sum of battery capacity become 90%.
Capacity (%)
≤ 30%
Alert message duration 5 sec.
2 Beep sound duration 0.5 sec.
≤ 15%
Shutdown Pop message
3 Beep sound duration 0.5 sec.
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
Event
30 seconds waiting
Shutdown
5831612-100 Revision 4
Note