Query Retrieve; Worklist; Report - GE Voluson Swift Instructions For Use Manual

Table of Contents

Advertisement

System Setup

QUERY RETRIEVE

WORKLIST

Private Tags
Modality
Add local data
ViewPoint Default
Note

REPORT

Note
DICOM Queue Status
Note
Note
Controls
Hold Queue
Retry
11-30
Default Appl. from the drop down menu.
Select the
The drop down menu contains exam applications available in patient dialog (Abdomen, Small
Parts, Obstetrics, Cardiology, Transrectal, Vascular, Gynecology, Pediatrics, Cephalic, MSK).
The selected exam application is used for all exams that are imported into the local archive
from a remote query/retrieve server.
Determines whether the private tags defined for communication with the Viewpoint -
worklist are used when querying the worklist.
Select either
"all".
yes : Locally stored patient data and patient data from the worklist are merged.
(Fields that are available in the worklist are taken from the worklist, fields that
are only available in the data base are taken from the local data base.)
no : Data only contained in the worklist is used to populate the patient data
fields. No locally stored data is used.
ask : A dialog is shown whenever there is data from the worklist and from the
local database available. Depending on the selection in the dialog, either the
action described under
Loads the default settings for ViewPoint.
Private Tags only work if the other system also supports Private Tags.
Select one of the following
Network : Send the patient report to a PC report station via DICOM network.
Serial : Send the patient report to a PC report station that is connected by serial port. The
optional "PRY USB-RS232 Connection kit" must be connected to the system.
When this Transfer Mode is selected, different fields are available:
control and Bits per second .
The bit rate (Bits per second) must be the same as on the receiving PC report station.
The Queue Status window displays all DICOM transfers which have not been sent, are being
sent at the moment or failed. (Successful transfers are deleted from the list). Status
displayed while the data is converted in the format that shall be sent later on (e.g. Converting
Raw Data to DICOM Multiframe).
If more than 600 entries exist, a message appears asking to clear the DICOM Queue. If the
number of 1500 entries is exceeded, the DICOM transfer is stopped completely due to
overflow. Again a message appears informing that no more transfer is possible and that the
DICOM Queue has to be cleared.
If the transfer was successful, but a storage commitment request was not yet successful, the
images receive the status sent. As soon as the storage commitment was successful the
entries (both images and storage commit) are deleted from the list.
The system no longer tries to send data in the queue. As soon as
pressed, the system continues to send data.
Retry the selected exams.
All or ULTRASOUND . No selection is also possible and defaults to
yes or no is executed.
Transfer Modes.
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
COM Port , Flow
Conv is
Process Queue is
5831612-100 Revision 4