Power Supply; Ac Power Cord Set; Opio; Console Design - GE Voluson Swift Instructions For Use Manual

Table of Contents

Advertisement

Technical Data/ Information

13.2.2 Power Supply

Power requirements:
Power consumption:
Thermal output:
Fan:
Fuse:

13.2.3 AC Power Cord Set

Power cord:

13.2.4 OPIO

OPIO:
Hard key buttons:
Recording keys:

13.2.5 Console Design

Probe ports:
Probe holder:
Solid State drive:
Peripherals:
Wheels:
Cables:
Handles:
13-4
Protection class I
100 - 240 VAC
Frequency: 50 Hz, 60 Hz (± 2%)
Max. 450VA including all options
Typical power consumption with ~400VA load approx. 1.75 A at 230V/50 Hz without
peripherals.
Max. 1370 BTU/h
Perceived noise level: max. 40 dB/A
Part number : 0034.1526
Technical spec:
Rated voltage : 250VAC
Rated Current : 10A
Breaking Capacity : 10A @250VAC
Operation speed : 40ms
Type SJT, SJE, SJO or SJTO, 14AWG, 3-Conductor,VW-1, 125V, 15A or 250 V, 10 A, max 4.0
m long;
One end with Hospital Grade Type, NEMA 5-15P or 6-15P. Other end with appliance coupler.
"CAUTION Grounding reliability can only be achieved when the equipment is connected to an
equipment receptacle marked "Hospital Only" or "Hospital Grade"".
Height adjustable:
Voluson SWIFT(18.5 inch) : 830mm ~ 1030mm
Voluson SWIFT+(21.5 inch) : 830mm ~ 1030mm
Rotation : Adjustable +/- 70° from center
Ergonomic layout, Interactive back-lightning
Integrated for remote control of up to 3 peripherals or DICOM devices, one dedicated
DVD recording key
3 active ports
Probe connection area designed to allow free movement of legs
5
Integrated SSD (Voluson SWIFT : 512 GB or 1TB, Voluson SWIFT+: 1TB)
On-board storage for peripherals: Black/white printer
Wheel diameter 125mm, integrated locking mechanism that provides rolling lock
Integrated cable management
Rear handle
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 Revision 4

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Voluson swift+

Table of Contents