Sealed Lead Acid Battery; System Switch - GE 9100c Technical Reference Manual

Anesthesia machine
Table of Contents

Advertisement

2.7.3 Sealed Lead
Acid Battery

2.7.4 System switch

M1207026
M1207027
Reproduced from the electronic master in MATRIX
A sealed lead acid battery supplies battery backup for the 9100c
ventilator.
The battery provides:
capacity to operate ventilator system for 30 minutes (fully
charged);
the battery is internally fused (auto-resettable).
Input: Normally 15 VDC at 25 °C during charge.
Output: +0.3 to 2 Amps during discharge.
The system switch has two positions: On and Standby.
In the Standby position
The switch turns off the ventilator (electrical).
In the On position
The switch turns on the ventilator (electrical).
004
MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English
2 Theory of Operation
2-21