Electrode Mark
N or RF

12.5.3 ECG Cine

12.5.3.1 ECG 2D Auto Cine

12.5.3.2 ECG Cine-Split Function

12.5.4 Safety Rules to be followed

Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 Revision 4
Remarks:
On the screen the ECG curve starts running from left to right if scan mode is active.
The most recent information is always on the right edge of the image.
ECG speed adjustment is only possible in scan mode.
ECG Color Code Cable
IEC
Color Code
R
Red
L
Yellow
Black
In the ECG memory a longer period than the one displayed on the monitor is stored. With the
help of the [Auto Cine] key the previous ECG curve can be scrolled back.
1. Use Format to change to the next (part of) frozen image sequence to play back the ECG
Cine memory.
2. Adjust the first trigger image with the trackball.
3. Switch the image position (press key again) and adjust the second trigger image with the
trackball.
Remark:
The green ECG line indicates to which image the trigger mark is related.
The Cine-Split Function is also possible in Auto Cine.
The ECG preamplifier is an integral part of the ultrasound scanner unit. The system may
only be operated in places that go conform with the rules for medically used locations.
The power cable of the ultrasound scanner system must not be connected to a damaged
socket. The socket must be equipped with a grounded conductor. If necessary a
potential equilibrium must be connected.
Only the patient cable provided by GE ULTRASOUND KOREA, LTD. may be used.
Consequently, only push-button electrodes may be used.
Take care that neither bare parts of one of the electrodes nor the patient can get in
contact with conductive parts (e.g., metal parts of the examination bed, trolley, or
similar).
This device must not be used for an intra-operative procedure on the heart.
If the use of a HF surgical unit with simultaneous connected ECG electrodes becomes
necessary, a maximum distance of ECG-electrodes from the surgical field and a correct
position and contact of the neutral electrode of the HF surgical unit must be observed
(avoidance of burning risk).
Note that stimulation current devices can influence the ECG signal.
AHA
Electrode Mark
Color Code
RA
White
LA
Black
RL
Green
Peripheral Devices
Position of the human body
surface
Right Arm
Left Arm
Right Leg (Neutral)
12-9