2.15 Software Update

Note
Info

Software download and installation

2.16 System messages

High system temperature

Note
2-30
Software update for the unit may become available for download and installation through the
GE Service platform. When a software update is available a message icon is displayed on the
status bar.
Software update through the GE service platform may not be available in all markets.
Please contact a service technician for further assistance.
1.
Press the standby button on the user interface.
The Shut Down dialogue window with the software
Download .
2.
Press
The download process is started. The progression of the download process is displayed.
Download may be paused and can be resumed later.
3.
When the software download is completed, the
Software installation may take about one hour.
Install .
4.
Press
The system reboots automatically and the installation process is started. Do not interrupt
the installation. The system may reboot automatically several times during installation or
you may be prompted to restart the system.
When the installation is complete, the New Software Verification window is displayed.
5.
Perform a check for all features listed. Move the cursor over the feature name to get
information on how to check each feature.
OK the signature field is enabled.
If all features are
6.
Enter your signature and press
The system is ready.
If one feature gets "Failed", the user will be prompted to reload the original software.
In case the temperature of the system is higher than usual, the following messages will be
displayed on the monitor screen. You can select the language in which the message is
displayed.
Caution
High system temperature reached! Further use of the system is possible. If this message
persistently appears, please contact your service agent.
High system temperature reached! If the system temperature stays high, the system will
shut down within 10 minutes.
If the system needs to shut down, an additional message will be displayed in the message
area of the monitor screen:
High system temperature - system shutdown within [ ] minutes!
Download button is displayed.
Download button changes to Install .
Send .
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use