LOGIQ e/LOGIQ e Vet/LOGIQ i/ Vivid e Basic Service Manual Part Number: 5370626-100 Revision: 18...
Page 2
This manual covers the software versions for the products as below: LOGIQ e: R4.x.x, R5.x.x, R6.x.x, R7.x.x LOGIQ e Vet: R5.x.x, R6.x.x, R7.x.x VIVID e: R4.x.x, R5.x.x, R6.x.x LOGIQ i: R4.x.x, R5.x.x...
Page 3
• VERSUCHEN SIE NICHT, DAS GERä T ZU REPARIEREN, BEVOR DIESES WARNUNG KUNDENDIENST-HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN (DE) WURDE. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLä GE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
Page 4
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual ESTE MANUAL DE SERVICIO Só LO EXISTE EN INGLé S. • SI ALGú N PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA QUE NO SEA EL INGLé S, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN SERVICIO DE TRADUCCIó...
Page 5
ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝ∆ΥΝΟΥΣ. EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ EL. • HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT, AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE. • NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A FIGYELMEZTETÉS KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
Page 6
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU. • EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN ENSKU, ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU. • REYNIÐ EKKI AÐ ÞJÓNUSTA TÆKIÐ NEMA EFTIR AÐ HAFA SKOÐAÐ OG VIÐVÖRUN SKILIÐ...
Page 7
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual šĪ APKALPES ROKASGRĀMATA IR PIEEJAMA TIKAI ANGĻU VALODĀ. • JA KLIENTA APKALPES SNIEDZĒJAM NEPIECIEŠAMA INFORMĀCIJA CITĀ VALODĀ, NEVIS ANGĻU, KLIENTA PIENĀKUMS IR NODROŠINĀT TULKOŠANU. • NEVEICIET APRĪKOJUMA APKALPI BEZ APKALPES ROKASGRĀMATAS BRĪDINĀJUMS IZLASĪŠANAS UN SAPRAŠANAS.
Page 8
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual ACEST MANUAL DE SERVICE ESTE DISPONIBIL NUMAI ÎN LIMBA ENGLEZĂ. • DACĂ UN FURNIZOR DE SERVICII PENTRU CLIENŢI NECESITĂ O ALTĂ LIMBĂ DECÂT CEA ENGLEZĂ, ESTE DE DATORIA CLIENTULUI SĂ FURNIZEZE O TRADUCERE.
Page 9
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual TA SERVISNI PRIROČNIK JE NA VOLJO SAMO V ANGLEŠČINI. • ČE PONUDNIK SERVISNIH STORITEV ZA STRANKO POTREBUJE NAVODILA V DRUGEM JEZIKU, JE ZA PREVOD ODGOVORNA STRANKA SAMA. •...
Page 10
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual BU SERVİS KILAVUZU YALNIZCA İNGİLİZCE OLARAK SAĞLANMIŞTIR. • EĞER MÜŞTERİ TEKNİSYENİ KILAVUZUN İNGİLİZCE DIŞINDAKİ BİR DİLDE OLMASINI İSTERSE, KILAVUZU TERCÜME ETTİRMEK MÜŞTERİNİN SORUMLULUĞUNDADIR. DİKKAT • SERVİS KILAVUZUNU OKUYUP ANLAMADAN EKİPMANLARA MÜDAHALE (TR) ETMEYİNİZ.
Page 11
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual (ZH-CN) (KO)
Page 12
GE Healthcare personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
Page 13
The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE Healthcare. GE Healthcare may revise this publication from time to time without written notice. TRADEMARKS All products and their name brands are trademarks of their respective holders.
Page 14
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Revision History Revision Date Reason for change 23, Oct. 2009 Initial Release 13, May 2010 Update for Software update and adding new probes 29, June 2010 Update to add Isolation Cart enhanced Version 30, November 2010 Update LOGIQ e Spare Parts and add USA CKD address 25, April 2011...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table of Contents CHAPTER 1 Introduction Overview ............1 - 1 Purpose of Chapter 1 .
Page 16
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual CHAPTER 2 Site preparations Overview............2 - 1 Purpose of this chapter 2 .
Page 17
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual CHAPTER 3 System Setup Overview............3 - 1 Purpose of Chapter 3 .
Page 18
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual CHAPTER 4 Functional Checks Overview............4 - 1 Purpose for Chapter 4 .
Page 19
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual CHAPTER 5 Components and Functions (Theory) Overview............5 - 1 Block Diagrams and Theory .
Page 20
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual CHAPTER 6 Service Adjustments Overview............6 - 1 Purpose of this chapter 6 .
Page 21
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual CHAPTER 7 Diagnostics/Troubleshooting Overview............7 - 1 Purpose of Chapter 7 .
Page 22
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual CHAPTER 8 Replacement Procedures Overview ............8 - 1 Purpose of Chapter 8 .
Page 23
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual CHAPTER 9 Renewal Parts Overview............9 - 1 Purpose of Chapter 9 .
Page 24
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual CHAPTER 10 Care & Maintenance Overview ............10 - 1 Periodic Maintenance Inspections .
Direction 5370626-100, Revision18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Chapter 1 Introduction Section 1-1 Overview 1-1-1 Purpose of Chapter 1 This chapter describes important issues related to safely servicing this ultrasound machine. The service provider must read and understand all the information presented here before installing or servicing a unit.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 1-1-4 Typical Users of the Basic Service Manual • Service Personnel (installation, maintenance, etc.). • Hospital’s Service Personnel • Contractors (Some parts of Chapter 2 - Site Preparations)
Page 27
Direction 5370626-100, Revision18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 1-1-5 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Models Covered by this Manual Table 1-2 LOGIQ e R4.x.x Model Designations Part Number Description 5151219 LOGIQ e R4.x.x Console for USA 5151220 LOGIQ e R4.x.x Console for Argentina 5151243...
Page 28
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 1-5 LOGIQ e R6.x.x Model Designations Part Number Description 5389812 LOGIQ e R6.x.x Console for Japanese 5389813 LOGIQ e R6.x.x Console for Australia 5389814 LOGIQ e R6.x.x Console for India 5410642 LOGIQ e R6.x.x Console for CKD USA 5410642...
Page 29
Direction 5370626-100, Revision18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 1-9 Vivid e R4.x.x Model Designations Part Number Description 5171613 Vivid e R4.x.x Console for USA 5172528 Vivid e R4.x.x Console for Argentina 5171633 Vivid e R4.x.x Console for Europe 5171603 Vivid e R4.x.xConsole for China 5171755...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 1-11 Vivid e R6.x.x Model Designations Part Number Description 5435033 Vivid e R6.x.x Console for USA 5435034 Vivid e R6.x.x Console for EU 5435035 Vivid e R6.x.x Console for Argentina 5435036 Vivid e R6.x.x Console for India 5435037...
Direction 5370626-100, Revision18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 1-2 Important Conventions 1-2-1 Conventions Used in Book Icons Pictures, or icons, are used wherever they reinforce the printed message. The icons, labels and conventions used on the product and in the service information are described in this chapter. Safety Precaution Messages Various levels of safety precaution messages may be found on the equipment and in the service information.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 1-2-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
Direction 5370626-100, Revision18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 1-2-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment. Table 1-16 Warnings LABEL/SYMBOL PURPOSE/MEANING LOCATION Identification and Rating Plate •...
Page 34
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 1-16 Warnings LABEL/SYMBOL PURPOSE/MEANING LOCATION “Protective Earth” indicates the Inside of AC adapter protective earth (grounding) terminal. “ON” indicates the power on position of the power switch. CAUTION Stick to Power Switch This Power Switch DOES NOT...
Page 35
Direction 5370626-100, Revision18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 1-16 Warnings LABEL/SYMBOL PURPOSE/MEANING LOCATION “Consult accompanying documents” is intended to alert the user to refer to the operator manual or other instructions Various when complete information cannot be provided on the label.
Page 36
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 1-16 Warnings LABEL/SYMBOL PURPOSE/MEANING LOCATION The separate collection symbol is affixed to a battery, or its packaging, to advise you that the battery must be recycled or disposed of in accordance with local or country laws.
Direction 5370626-100, Revision18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 1-3 Safety Considerations 1-3-1 Introduction The following safety precautions must be observed during all phases of operation, service and repair of this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this manual, violates safety standards of design, manufacture and intended use of the equipment.
If the battery emits an odor or heat, is deformed or discolored, or in a way appears abnormal during use, recharging or storage, immediately remove it and stop using it. If you have any questions about the battery, consult GE or your local representative. •...
Direction 5370626-100, Revision18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 1-3-7 Dangerous Procedure Warnings Warnings, such as the examples below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed. DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT DANGER IN THIS EQUIPMENT.
GE is not responsible for the loss of this data. If PHI (Patient Healthcare Information) data needs to be sent to GE employees for service purposes, GE will ascertain agreement from the customer. The patient information shall only be transferred by...
Direction 5370626-100, Revision18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 1-4 EMC, EMI, and ESD 1-4-1 Electromagnetic Compatibility (EMC) Electromagnetic compatibility describes a level of performance of a device within its electromagnetic environment. This environment consists of the device itself and its surroundings including other equipment, power sources and persons with which the device must interface.
Canada 1-800-668-0732 Service 1-800-321-7937 Latin America Applications support 1-262-524-5698 Europe GE Ultraschall Deutschland GmbH& Co. KG Phone: +33 (0)130-831-300 (General Imaging and Cardiac) BeethovenstraBe 239 Postfach 11 05 60, D-42665 Solingen Fax: +49 (0)212-2802-431 Germany Asia (Singapore) Tel: +65 291-8528...
P.R.China 214028 GE Ultrasound Korea: TEL: +82-31-740-6112 65-1, Sangdaewon-dong, Jungwon-gu, Seongnam-si, Gyeonggi-do, FAX: +82-31-740-6435 Korea 462-120 GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC DBA GE Healthcare TEL: (1) 800-437-1171 9900 Innovation Drive FAX: (1) 414-721-3865 Wauwatosa, WI 53226 U.S.A...
Page 44
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual This page was intentionally left blank. 1-20 Section 1-5 - Customer Assistance...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Chapter 2 Site preparations Section 2-1 Overview 2-1-1 Purpose of this chapter 2 This chapter provides the information required to plan and prepare for the installation of a LOGIQ e/ LOGIQ e Vet/LOGIQ i/Vivid e .
2-2-2 Electrical Requirements NOTE: GE Medical Systems requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
Page 47
2-2-2-5 Unit Power Plug If the unit arrives without a power plug, or with the wrong plug, you must contact your GE dealer or the installation engineer must supply what is locally required. 2-2-2-6 Power Stability Requirements Voltage drop-out Max 10 ms.
Never place a label where RF gaskets meet the unit. Otherwise, the gap created will permit gaskets touch metal RF leakage. Or, if a label has been found in such a position, move the label. Use GE specified harnesses The interconnect cables are grounded and require ferrite beads and other shielding. Also, and peripherals cable length, material, and routing are all important;...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 2-2-4 Scan Probe Environmental Requirements Operation:10° to 40° C Storage:-10° to 50° C NOTE: The recommended storage temperature range of 6Tc-RS is between 0°C and 45°C. NOTE: Temperature in degrees C.
All electrical work on these product must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment.
2-3-2 Required Features NOTE: GE Medical Systems requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
Page 52
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 2-3-3-1 Recommended and Alternate Ultrasound Room Layout Recommended standard floor plan and a minimal floor plan for ultrasound equipment: 18 IN . LINEN SUPPLY (46 C M) DE DIC AT E D P O WE R R E C EP TA C LE...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 2-3-4 Networking Pre-installation Requirements 2-3-4-1 Stand Alone Scanner (without Network Connection) None. 2-3-4-2 Scanner Connected to Hospital’s Network Supported networks: Wireless LAN 2-3-4-3 Purpose of DICOM Network Function DICOM services provide the operator with clinically useful features for moving images and patient information over a hospital network.
Page 54
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 2-3-4-4 DICOM Option Pre-installation Requirements (cont’d) LOGIQ e/ IP Address Local Port Host Name AE Title Net Mask ROUTING INFORMATION GATEWAY IP Addresses Destination IP Addresses Default ROUTER1 ROUTER2 ROUTER3...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Chapter 3 System Setup Section 3-1 Overview 3-1-1 Purpose of Chapter 3 This chapter contains information needed to install the unit. Included are references to a procedure that describes how to receive and unpack the equipment and how to file a damage or loss claim.
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 3-1-2 Average Installation Time Table 3-2 Average Installation Time Description Average Installation Time Comments 20 minutes Unpacking the scanner 30 minutes Scanner wo/options Dependent on the configuration that is required 30 minutes DICOM Option Dependent on the amount of configuration...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 3-1-3 Installation Warnings 1.) There are no operator serviceable components. To prevent shock, do not remove any covers or panels. Should problems or malfunctions occur, unplug the power cord. Only qualified service personnel should carry out servicing.
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 3-2 Receiving and Unpacking the Equipment When a new system arrives, check that any components are not damaged and are not in short supply. If shipping damage or shortage occurs, contact the address shown in Chapter 1. 1.) Cut the four PLASTIC BANDs.
Page 59
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 3-2 Receiving and Unpacking the Equipment (cont’d) 3.) Take out the Paper pad. 4.) Take out console together with 2 interleavers from console package. 5.) Take out the interleavers beside Accessories Package. 6.) Take out Accessories Package.
Page 60
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 3-2 Receiving and Unpacking the Equipment (cont’d) 7.) Remove 2 interleavers. 8.) Remove plastic bag. Interleaver Console Interleaver Figure 3-3 Removing interleavers and plastic bag Section 3-2 - Receiving and Unpacking the Equipment...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 3-2 Receiving and Unpacking the Equipment (cont’d) NOTE: Check the shipping container for special instructions. Verify that the container is intact. In some cases a secondary container may be used. If so, ask the carrier for unpacking instructions. Figure 3-4 Labels on Package Please carefully unpack the system, and do not dispose the package of LOGIQ e/LOGIQ e Vet/ CAUTION...
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 3-4 Preparing for Installation 3-4-1 Verify Customer Order Compare items received by the customer to that which is listed on the delivery order. Report any items that are missing, back ordered or damaged.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 3-5 Completing the Installation 3-5-1 Power On / Boot Up NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly.
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 3-5-1-2 Turn on the system Press the Power On/Off switch at the front of the system once. Power On/Off switch Figure 3-6 Power On/Off Switch When the Power On/Off switch on the Control Panel is pressed once, the Back-end Processor starts and the software code is distributed to initiate the scanner.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 3-5-2-2 Scanner Shutdown Disconnect the Mains Power Cable if necessary. For example: Relocating the scanner. 3-5-3 Transducer Connection 1.) Carefully open the system LCD display, plug the probe connector into the probe port, then lock the probe latch upward.
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 3-6 System Configuration 3-6-1 System Specifications 3-6-1-1 Physical Dimensions and weight with old LCD The physical dimensions of the LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e console with old LCD are summarized in Figure 3-9 on page 3-12 Table 3-4...
Page 67
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 3-6-1-2 Physical Dimensions and weight with new LCD The physical dimensions of the LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e console with new LCD are summarized in Figure 3-10 on page 3-13 Table 3-5 Physical Dimensions of LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e with new LCD Height...
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 3-6-1-3 How to identify old LCD and new LCD • Identify LCD type from the appearance of Ultrasound System. old LCD new LCD old LCD new LCD new LCD old LCD Figure 3-11 Difference between Old LCD and New LCD...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 3-6-3 Approved peripherals Table 3-7 Approved peripherals Device Manufacturer Model Interface Remark SONY UP-D897 B/W Printer SONY UP-D898MD MITSUBISHI P95D SONY UP-D23MD Digital Color Printer SONY UP-D25MD HP Officejet Pro K5400 HP Officejet Pro K8600 HP Color Printer...
Page 70
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 3-7 Approved peripherals Device Manufacturer Model Interface Remark MD-800E Not supported by LOGIQ e R4.x.x, Vivid e DVD-Recorder Panasonic MD-800U R4.x.x, LOGIQ i R4.x.x Pinnacle Video Transfer - 8230- Not supported by LOGIQ e R4.x.x/5.0.x, 10022-11 Video Transfer...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 3-6-4 Connecting Cables Equipment damage possibility. Be sure to use the following recommended connecting cables to WARNING WARNING connect recording devices and a network with LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e console. Table 3-8 List of Connecting Cables Name...
Page 72
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 3-6-5-2 Left Panel Connector Located on the left side panel are two USB Ports for control connections for printer, service tools and earphone port. Figure 3-13 Left Side Panel 1.) Two interchangeable USB Port (Digital Printers, DVD-RW and/or FootSwitch, etc) 2.) Earphone Port.
Page 73
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 3-6-5-3 This section indicates the pin assignment for each connector. 1. Pin Assignment of DC input Connector: 4 Pin, Female Table 3-9 Pin Assignments of DC input Pin No.
Page 74
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 3-6-5-4 Connect peripherals A.) Connect B/W printer to the system. B/W Printer can be properly connected using USB Port1 or USB Port 2 (Figure 3-14 on page 3-20 Figure 3-14 Connect B/W printer to the system B.) Connect UP-D23MD color printer to the system.
Page 75
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 3-6-5-4 Connect peripherals (cont’d) C.) Connect DVD-RW to the system. DVD-RW can be properly connected using USB Port1 or USB Port2 (Figure 3-16 on page 3-21 Figure 3-16 Connect DVD-RW to the system DO NOT connect DVD-RW to the system while scanning.
Page 76
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 3-6-5-4 Connect peripherals (cont’d) D.) Connect FootSwitch to the system. FootSwitch can be properly connected using USB Port1 or USB Port2. 1-pedal footswitch 3-pedal footswitch Figure 3-17 Connect Foot Switch to the system E.) Connect Wireless LAN Adapter to the system.
Page 77
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 3-6-5-4 Connect peripherals (cont’d) G.) Connect the USB Memory to the system. The USB Memory can be properly connected using USB port 1 or 2. Figure 3-19 USB Memory Connection H.) Connect the ECG to the system.
Page 78
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 3-6-5-4 Connect peripherals (cont’d) I.) Connect the USB HDD to the system. The USB Harddisk can be properly connected using USB port 1 and 2. Figure 3-21 Emergency Disk Connection J.) Connect the BarCode Reader to the system, the BarCode Reader can be properly connected via USB port 1 or 2.
Page 79
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 3-6-5-4 Connect peripherals (cont’d) K.) Connect the Sandisk U3 USB Flash Drive to the system. The Sandisk U3 USB Flash Drive can be properly connected using USB port 1 and 2. Figure 3-23 Sandisk U3 USB Flash Drive Connection NOTE: The Password of the Sandisk U3 USB Flash Drive is Predefined.
Page 80
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 3-6-5-4 Connect peripherals (cont’d) Figure 3-25 Wrong password error Click OK, input the password again. The Sandisk U3 USB Flash Drive will be locked if reaching the maximum number of failed password attempts.
Page 81
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 3-6-5-4 Connect peripherals (cont’d) L.) Connect the external monitor to the system. The monitor can be properly connected using the VGA port. Figure 3-27 VGA Connection Press Ctrl+Alt+V on the keyboard, a dialog box appears. For LOGIQ e R5.x.x, R6.x.x, R7.x.x, LOGIQ i R5.x.x, For LOGIQ e R4.x.x, LOGIQ i R4.x.x, Vivid e R4.x.x Vivid e R5.x.x/R6.x.x, LOGIQ e Vet R6.x.x, R7.x.x...
Page 82
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 3-6-6 Available Probes See in specification in the LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e User Reference Manual for Probes and intended use. For LOGIQ e R4.0.x, the system supports 4C-RS, E8C-RS, 8C-RS, i12L-RS, 8L-RS, 3S-RS, 12L-RS Probes.
VALOX MICRO-CONVEX H41562LL 5134643 TRANSRECTAL VETERINARY 8C-RS Vet VALOX MICRO-CONVEX H41562LK 5134642 PEDIATRIC NEONATAL INTRAOPERATIVE SMALL i12L-RS Vet ABS (GE) LINEAR H41562LN 5134645 PARTS VASCULAR PEDIATRICS SMALL PARTS 8L-RS Vet VALOX LINEAR H41562LM 5134644 PERIPHERAL VASCULAR CARDIOLOGY 3S-RS Vet VALOX...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 3-8 Connectivity Installation Worksheet Site System Information Comments: Floor: Site: Dept: Room: LOGIQ SN: Type: REV: CONTACT INFORMATION Name Title Phone E-Mail Address TCP/IP Settings Name - AE Title: IP Settings Remote Archive Setup IP Address:...
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 3-9 Loading Base Image Software Please refer to: Section 8-3 "Loading the System Software" on page 8-4 3-32 Section 3-9 - Loading Base Image Software...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 3-10Software Version check out 3-10-1 Functional Check-out 1.) Power on LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e scanner and wait until system booting to main screen. 2.) Press Utility/Config key on control panel. 3.) Choose the About button on the right.
User Manual(s) User Check that the correct User Manual(s) for the system and software revision, is included with the installation. Specific language versions of the User Manual may also be available. Check with your GE Sales Representative for availability. 3-34...
Page 89
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose for Chapter 4 This chapter provides procedures for quickly checking major functions of the LOGIQ e/LOGIQ e Vet/ LOGIQ i/Vivid e console, diagnostics by using the built-in service software, and power supply adjustments.
Page 90
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 4-3 General Procedure SYSTEM REQUIRES ALL COVERS CAUTION Operate this unit only when all board covers and frame panels are securely in place. The covers are required for safe operation, good system performance and cooling purposes. Lockout/Tagout Requirements (For USA only) NOTICE Follow OSHA Lockout/Tagout requirements by ensuring you are in total control of the Power Cable on...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-1-1 Scanner Power On (cont’d) When power is applied to the scanner, power is distributed to the Cooling Unit, Control Panel, LCD, Peripherals and the Back-end Processor. 4-3-1-2 Turn on the system Press the Power On/Off switch at the front of the system once.
Page 92
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-2-1 Back-end Processor Power Down (cont’d) 4.) The shutdown process takes 15 seconds and the power off sequence is complete when the power status LED is turned off. 5.) Disconnect the probes.Clean or disinfect all probes as necessary.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-3 Archiving and Loading Presets NOTE: Always save presets before any software reload. This ensures the presets loaded after the software reload are as up–to–date as possible. All user presets except changes to Summary, Anatomy, and Biometry pages, can be saved on an DVD- R disk (or USB memory device) for reloading on the system.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-3-2 Loading Presets from an DVD-R disk (or USB memory device) 1.) Insert the DVD-R disk with the archived Presets into the DVD-RW. 2.) Access to Utility->System->Backup/Restore. The Restore sheet will be shown on the LCD display. (see Figure 4-3 on page 4-5 3.) Select the items needed to be stored.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-6 System Features 4-3-6-1 Control Panel Figure 4-4 Control Panel Tour 1.) TGC 2.) New Patient 3.) End Exam 4.) Mode/Gain/Auto Keys 5.) Preset Key 6.) Imaging/Measurement Keys 7.) Depth/Zoom/Ellipse 8.) Image Keys 9.) Print Keys...
Page 96
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-6-2 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e SoftMenu Key Tour In general, there are two types of softMenu keys: Paddle Switch and adjustable knobs. Figure 4-5 SoftMenu Key Tour 1.) The Paddle Switch is used to access and adjust the Sub SoftMenu.
Page 97
2. Patient Name, Patient Identification. 14. Focal Zone. 3. Acoustic Output Readout, 15. TGC (not shown on the image). 4. GE Symbol: Probe Orientation Marker. Coincides with a 16. Body Pattern. probe orientation marking on the probe. 5. Image Preview.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-7 B Mode Checks 4-3-7-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-28, in Chapter 3 Installation to the System probe connector. 2.) Turn ON the scanner (if it isn’t turned on already) NOTE: The keyboard layout varies from each Product, the following figures show LOGIQ e R5.x.x as...
Page 99
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-7-2 B Mode OP Panel Controls Table 4-3 B Mode Control Panel Controls Step Task Expected Result(s) Comments Press B Mode key B Mode Starts Adjust the field of view. Increasing the depth may view larger/deeper structures rates, and decreasing the depth may view near the skin Adjust Depth...
Page 100
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-7-3 B Mode Softmenu Key Table 4-4 B Mode Softmenu Key Step Task Expected Result(s) comments Selects a level below which echoes will not be Adjust Rejection amplified (an echo must have a certain minimum amplitude before it will be processed).
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-8 M Mode Controls 4-3-8-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-28, in Chapter 3 Installation to the System probe connector. 2.) Turn ON the scanner (if it isn’t turned on already).
Page 102
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-8-2 M Mode OP Panel Controls Table 4-5 M Mode OP Panel Controls Step Task Expected Result(s) Comments Press M Mode key M Mode Starts Controls the amount of echo information displayed in an image.
Page 103
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-8-3 M Mode Softmenu Key Table 4-6 M Mode Softmenu Key Step Task Expected Result(s) Comments Selects a level below which echoes will not be amplified (an echo must have a certain Adjust Rejection minimum amplitude before it will be processed).
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-9 Color Flow Mode Checks 4-3-9-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-28, in Chapter 3 Installation to the System probe connector. 2.) Turn ON the scanner (if it isn’t turned on already).
Page 105
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-9-2 Color Flow Mode OP Panel Controls Table 4-7 Color Flow Mode OP Panel Controls Step Task Expected Result(s) comments Press CFM-Mode key CFM Mode Starts Amplifies the overall strength of the echoes processed in the Color Flow window.
Page 106
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-9-3 Color Flow Mode Softemenu Key Table 4-8 Color Flow Mode Softmenu Key Step Task Expected Result(s) Comments Threshold assigns the gray scale level at Threshold which color information stops. Controls the number of samples gathered for Packet Size a single color flow vector.
Page 107
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 4-8 Color Flow Mode Softmenu Key Step Task Expected Result(s) Comments Velocity scale determines pulse repetition frequency. If the sample volume gate range Change PRF exceeds single gate PRF capability, the For LOGIQ e, LOGIQ system automatically switches to high PRF e Vet and LOGIQ i...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-10 Doppler Mode Checks 4-3-10-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-28, in Chapter 3 Installation to the System probe connector. 2.) Turn ON the scanner (if it isn’t turned on already).
Page 109
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-10-2 Doppler Mode OP Panel Controls Table 4-9 Doppler Mode OP Panel Controls Step Task Expected Result(s) Comments Press PW Mode key PW Mode Starts Amplifies the overall strength of the echoes processed in the Color Flow window.
Page 110
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-10-3 Doppler Mode OP Panel Controls Table 4-10 Doppler Mode Touch Panel Controls Step Task Expected Result(s) comments Selects a level below which echoes will not be amplified (an echo must have a certain Adjust Rejection minimum amplitude before it will be...
Page 111
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 4-10 Doppler Mode Touch Panel Controls Step Task Expected Result(s) comments Adjusts the baseline to accommodate faster Move Baseline or slower blood flows to eliminate aliasing. Velocity scale determines pulse repetition frequency.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-11 CWD Functional Check 4-3-11-1 Preparations 1.) Connect 3S-RS/6S-RS/6Tc-RS Probe to the system, see in Chapter 3 Installation. 2.) Turn ON the scanner (if it isn’t turned on already). 4-3-11-2 Activating CW Doppler To activate CW Doppler Mode:...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-12 Basic Measurements NOTE: In some systems, key MESAURE is marked as Calcs; Set is marked as Enter. NOTE: The following instructions assume that you first scan the patient and then press Freeze. 4-3-12-1 Distance and Tissue Depth Measurements 1.) Press MEASURE once;...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-13-2 Activating the probe The probe activates in the currently-selected operating mode. The probe's default settings for the mode and selected exam are used automatically. 4-3-13-3 Deactivating the probe 1.) Press the Freeze key.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-14-5 Moving through a CINE Loop Frame By Frame Adjust the Frame by Frame up/down Two-Button Softkey to move through CINE memory one frame at a time. 4-3-15 Image Management (QG) For Image Management functionality refer to the LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Quick Guide.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-16 Backup and Restore Database, Preset Configurations and Images 4-3-16-1 Formatting Media 1.) To format the backup media, DVD-RW, select the UTILITY/CONFIG button on the Keyboard. 2.) Select CONNECTIVITY, then REMOVABLE MEDIA. Properly label and Insert the backup media.
Page 117
4.) On the LCD display, select BACKUP/RESTORE. NOTE: If you are not logged in as GE Service or with administrator privileges, the Operator Login window is displayed. Log on with administrator privileges. 5.) In the Backup list, select Patient Archive, Report Archive and User Defined Configuration.
Page 118
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-16-3 Restore System Presets and Configurations The restore procedure overwrites the existing database on the local hard disk drive. Make sure to CAUTION insert the correct DVD (or USB memory device). 1.) Insert the Backup/Restore DVD (or USB memory device) into the drive.
Page 119
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-16-4 Archiving Images 1.) Insert the archive media. To format the archive media, DVD-RW, select the Utility/Config button on the Keyboard. NOTE: For LOGIQ e R5.x.x, LOGIQ e R6.x.x, LOGIQ e R7.x.x, LOGIQ e Vet, Vivid e R5.x.x, Vivid e R6.x.x and LOGIQ i R5.x.x, USB HDD also can be used as archive media.
Page 120
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-16-4 Archiving Images (cont’d) 7.) Press PATIENT and set the Dataflow to store images directly to DVD-ROM. 8.) From the image screen, press EZBACKUP/MOVE. The Move Images pop-up appears. Figure 4-21 Image Archive Move Pop-up Menu 9.) Press Next on pop-up message.
Page 121
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-16-4 Archiving Images (cont’d) 11.)Press Finish after Backup/Move complete. Figure 4-23 EZBackup/Move complete All databases, presets and images should now be saved to removable media. 4-3-16-5 Emergency Repair Disk 1.) Connect the USB HDD to system.
Page 122
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-16-5 Emergency Repair Disk (cont’d) 4.) Return to scan mode. Press Ctrl + B, the Emergency Disk Making window pop-up appears. Figure 4-25 Emergency Disk Making Pop-up Menu 5.) Press Next on pop-up message.
Page 123
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-16-5 Emergency Repair Disk (cont’d) 7.) Press Finish after Emergency Disk Making complete. Figure 4-27 Emergency Disk Making Complete All databases, presets and images should now be saved to Emergency Disk. 8.) Press Ctrl + R, Emergency Disk Recovery will pop-up.
Page 124
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 4-3-16-5 Emergency Repair Disk (cont’d) 10.)A recovery information message appears. Press Next to continue. Figure 4-29 Recovery Information message 11.)Press Finish after Emergency Disk Recovery complete. Figure 4-30 Emergency Disk Recovery Complete 4-36 Section 4-3 - General Procedure...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 4-4 Software Configuration Checks Table 4-11 Software Configuration Checks Step Task to do Expected Result(s) Check Date and Time setting Date and Time are correct Check that Location (Hospital Name) is correct Location Name is correct Check Language settings Desired Language is displayed...
Page 126
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual This page was intentionally left blank. 4-38 Section 4-5 - Peripheral Checks...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Chapter 5 Components and Functions (Theory) Section 5-1 Overview This chapter explains LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e ’s system concepts, component arrangement, and subsystem function. It also describes the Power Distribution System (PDS) and probes.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 5-2 Block Diagrams and Theory 5-2-1 Block Diagram TX64 15” XGA TMST(RFI/Carrier) Speaker RX64 LV unit On board Distributed Peripherals HV unit (option) Power DVD-RW Keyboard USB-Wireless A/N key USB2.0...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 5-2-2 General Information • LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e is an ultrasound imaging scanner. • The system can be used for: 2D Black and White imaging 2D Color Flow M-Mode Black and White imaging Doppler...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 5-2-3 External I/O Figure 5-2 External I/O module overview External I/O: The external I/O is the interface between the scanner and all external items. Examples: wireless network, USB interface medical grade printer and external medical grade SVGA displayer. 5-2-4 Peripherals DVD-RW, and a Black &...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 5-3 Power Diagrams 5-3-1 Overview The AC Power assy’s main tasks are to isolate and output to the DC/DC unit which is inside the system console. The input of AC power pack will be the AC outlet and it’s universal, the range is AC 90V-264V, 47-63Hz.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 5-3-3 Battery charging The charging circuit is lithium-Ion battery charge and discharge controller. This block can switch the power between the battery and the output of AC Pack. If the output of AC Pack is available, the power input of Charge Board Unit should be from the AC Pack and the battery will be charged if it’s not full.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 5-3-4 Air Flow Distribution Figure 5-5 Air Flow Inside the Scanner The two air flow passes allow the scanner to be cooled down as shown in the figure above. •...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 5-3-5 Fans Figure 5-6 Left & Right Fans The scanner contains the four fans at the following positions for producing an air flow. One fan assy: On the Left side for air flow path A. One fan assy: On the Right side for air flow path B.
Global Service User Interface (GSUI) 5-4-2-1 Internationalization The user interface provided by the service platform is designed for GE personnel and as such is in English only. There is no multi-lingual capability built into the Service Interface. 5-4-2-2 Service Login Select Utility/Config->Service.
Page 137
Password Operator Authorized access to specified diagnostics, error logs and utilities. Same Administrator acquisition diagnostic tests as GE Service. External Service gogems Every access request, whether successful or not, will be logged into a service access log that is viewable to authorized users.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 5-4-3 Service Home Page The navigation bar at the top of the screen allows the user to select from several tools and utilities. Figure 5-10 Customer Service Home Page 5-12 Section 5-4 - Common Service Platform...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 5-4-4 Error Logs Tab From the Error Logs Tab the Log Viewer displays four categories with pull-down sub-menus and an Exit selection. The Service Interface allows scanner logs to be viewed by all service users. The log entries are color-coded to identify the error level severity at a glance.
Page 140
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 5-4-4-1 Logs The seven sub-menus of the Logs category are System, Power, Infomatics, Temperature, Probe, Board, and DICOM. NOTE: Figure 5-11 provides a graphical example of the log entries for the System Logs. Log table headings for the different logs are as follows: •...
Page 141
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 5-4-4-2 Utilities The two sub-menus of the Utilities category are Plot Log, and Plot Page. Plot Log Plot Page Figure 5-12 Utilities Sub-Menus • Plot Log Allows for the color coded plot of all Log contents with the package on the ‘x’ axis and incident count on the ‘y’...
Page 142
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 5-4-4-4 Exit The sub-menu, Exit Log Viewer, returns the user to the Service Desktop home page. Figure 5-14 Exit Log Sub-Menu 5-16 Section 5-4 - Common Service Platform...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 5-4-5 Diagnostics Detailed Diagnostic information is found in Chapter Figure 5-15 User Diagnostic Page 5-4-5-1 Diagnostics Execution Diagnostic tests are executable by both local and remote users. The Service Platform provides top-level diagnostic selection based on the user’s level and login access permissions.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 5-4-6 Image Quality The Image Quality page is intended to contain tools for troubleshooting image quality issues. FIELD IS NOT YET POPULATED Figure 5-16 Image Quality Page 5-4-7 Calibration The Calibration page is intended to contain the tools used to calibrate the system.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 5-4-8 Configuration The Configuration page is intended to be used to setup various configuration files on the system. The Service Platform is the access and authorization control for remote access to the configuration subsystem.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 5-4-10 Replacement The Replacement page is intended to contain the tools used to track replacement parts used in the system. FIELD IS NOT YET POPULATED Figure 5-19 Part Replacement Page 5-4-11 The PM page is intended to contain the tools used in periodic maintenance of the system.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of this chapter 6 This section describes how to test and adjust the scanner. These tests are optional. You may use them to check the system for errors.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 6-2 Monitor Adjustments 6-2-1 Adjustments Procedures To adjust the brightness and volume: 1.) Press Ctrl + brightness key together (up/down), located on the keyboard. 1.) Press Ctrl + volume key together (left/right), located on the keyboard. Figure 6-1 LCD Monitor Section 6-2 - Monitor Adjustments...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter 7 This section describes how to setup and run the tools and software that help maintain image quality and system operation.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 7-2 Gathering Trouble Data 7-2-1 Overview There may be a time when it would be advantageous to capture trouble images and system data (logs) for acquisition to be sent back to the manufacturer for analysis. There are different options to acquire this data that would give different results.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-2-3 Collect a Trouble Image with Logs If the system should malfunction, press the Alt-D keys simultaneously. This will collect a screen capture of the image monitor, system presets and the following logs: Keyboard Shadow Log Error Logs Crash Log...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 7-3 USB Quick Save 7-3-1 Overview There may be times when the customer or field engineer will want to directly save images into USB memory. This is accomplished by saving individual Cine clips (moving images- avi format) or still images (jpg format) directly to a USB memory disk by pressing a Print Key.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-3-2 Check and Record the P3 Key Function (cont’d) 5.) In the Destinations section, record the service that is displayed. 6.) In the Physical Print Buttons section, record the parameters related to the service. 7-3-3 Setting the P3 Key to USB Quick Save If the P3 Key is not set to USB Quick Save:...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 7-4 Screen Captures There may be times when the customer or field engineer will want to capture a presentation on the screen. This is accomplished by first saving the image(s) to the clipboard using a Print Key. There’s no factory default print key to accomplish a secondary screen capture.
Page 155
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-4-3 Capturing a Screen (cont’d) Clipboard Thumbnai Menu Figure 7-3 Select Image to Capture 3.) Click FREEZE to unfreeze the image to view the image screen and the snapshots displayed on the bottom.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-4-3 Capturing a Screen (cont’d) For LOGIQ e R5.x.x, R6.x.x, R7.x.x, LOGIQ i R5.x.x, Vivid e R5.x.x, For LOGIQ e R4.x.x, LOGIQ i R4.x.x, Vivid e R4.x.x R6.x.x, LOGIQ e Vet R6.x.x, R7.x.x Figure 7-5 Save Dialog Box 6.) A Save dialog box will be opened.
1.) Press Utility/Config - Service, see Figure 7-6 on page 7-9 Figure 7-6 system Service button 1.) Select GE Service in option of User Level of GEMS Service Home page and input correct password, press OK button. See Figure 7-7 on page 7-9...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-5-2 Active Diagnostic Function 2.) Choose Diagnostic in Global Service Interface to active diagnostic functions, choose options to activate various functions correspondingly. Figure 7-8 Active Diagnostic Function 7-5-3 Control Frame Contains the user interface elements used for:...
Page 159
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-5-3-4 Short Text Message Displays either a starting message or aborting message, as well as the diagnostic completion status. 7-5-3-5 Background Color Initially gray, the Control Frame background color changes upon completion of a diagnostic to indicate completion status.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 7-6 Common Diagnostics 7-6-1 Utilities Provides two selections: 7-6-1-1 Disruptive Mode Allows you to enable or disable disruptive mode troubleshooting. 7-6-1-2 System Shutdown Allows for system shutdown from the diagnostic menu. Select to Restart System or Shutdown System. Also, select to retain Disruptive Mode or Not.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-6-2 PC Diagnostics (Non-Interactive Tests) 7-6-2-1 Essential Tests 7-6-2-2 System Board 7-6-2-3 Hard Disk Drive Surface Scan 7-6-2-4 Hard Disk Drive Quick Test 7-6-2-5 PCIBus 7-6-2-6 Video Card 7-6-2-7 Network Interface 7-6-2-8...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-6-3 PC Diagnostics (Interactive Tests) 7-6-3-1 Keyboard Test 7-6-3-2 Mouse Test 7-6-3-3 Audio Card Test (only for LOGIQ e R5.2.x, R6.x.x, R7.x.x, LOGIQ e Vet R6.x.x, R7.x.x) 7-6-4 Restart the system after diagnostics Always shutdown the system and reboot after a diagnostics session.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 7-7 Network and Insite II Configuration 7-7-1 Network Configuration 7-7-1-1 Wire-LAN Network 1.) Connect system with network. 2.) Press Utility/Config -> Connectivity -> TCPIP, in IP Settings window check the Enable DHCP box. In Network Speed box, choose the proper speed available.
Page 164
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-7-1-1 Wire-LAN Network (cont’d) NOTE: If user want to setup static IP address, uncheck Enable DHCP option, input static address in IP-Address box, Subnet Mask and Default Gateway box. In Network Speed box, choose the proper speed available.
Page 165
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-7-1-2 Wireless LAN Network 1.) Connect wireless LAN card with system. 2.) Click Start at the left bottom of screen, and select “Network Connections. Figure 7-13 Start menu 3.) In the pop-up window choose Wireless Network Connection.
Page 166
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-7-1-2 Wireless LAN Network (cont’d) 4.) The Wireless Network Connection dialog box displays. Select the proper wireless network in Available wireless networks window, check the Allow me to connect... box, and click Connect. For LOGIQ e R4.x.x, LOGIQ i R4.x.x, Vivid e R4.x.x For LOGIQ e R5.x.x/R6.x.x/R7.x.x, Vivid e R5.x.x/R6.x.x LOGIQ i R5.x.x, LOGIQ e Vet R6.x.x/R7.x.x...
SAVE. Figure 7-18 Advanced_Configuration page (out of GE network) NOTE: If system is within GE network, choose Development as server type, and do not need step 9.) enter Proxy Server address. Figure 7-19 Advanced_Configuration page (within GE network)
Page 168
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-7-2 Insite II Configuration (cont’d) 8.) Change to Device Information page, enter Device and Serial Number as the serial number of that system, then press Modify Device Configurations. Figure 7-20 Device_Information page NOTE: The Device name and Serial Number entered should be unique on server.
Page 169
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-7-2 Insite II Configuration (cont’d) 9.) Change to Proxy_Configuration page, enter Proxy Server Address as 3.20.128.6 and Proxy Server Port as 88, then press SAVE. Figure 7-22 Proxy_Configuration page Chapter 7 Diagnostics/Troubleshooting 7-21...
Page 170
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-7-2 Insite II Configuration (cont’d) 10.)Change to Diagnostics page. Enable Disruptive Mode at Service side by pressing Yes. Figure 7-23 Enable Disruptive Mode 11.)Customer will receive a request of Service side, click Yes to accept. Figure 7-24 Accept requirement of Service 7-22 Section 7-7 - Network and Insite II Configuration...
Page 171
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-7-2 Insite II Configuration (cont’d) 12.)Service can do Remote Diagnostics successfully. Figure 7-25 Remote Diagnostic successful 13.)Press power switch, and choose Full Maintenance Reboot to reboot system. Figure 7-26 Full Maintenance Reboot window Chapter 7 Diagnostics/Troubleshooting 7-23...
Page 172
Insite II Configuration (For LOGIQ e R6.x.x/R7.x.x and LOGIQ e Vet R6.x.x/ R7.x.x) 1.) Enter global service user interface, and log in as GE Service. See 7-5-1. 2.) On Configuration page, select Insite ExC Agent Configuration in the left column, fill in the information in Agent Configuration, choose “PPODUCT”...
Page 173
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-7-3 Insite II Configuration (For LOGIQ e R6.x.x/R7.x.x and LOGIQ e Vet R6.x.x/ R7.x.x) (cont’d) 3.) Set Disruptive Mode Status as Enabled. Change to Utilities page. select Disruptive Mode in the left column, Check whether the Disruptive Mode Status is Enabled, if not, click Yes to change the Disruptive Mode Status as Enabled.
Page 174
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-7-3 Insite II Configuration (For LOGIQ e R6.x.x/R7.x.x and LOGIQ e Vet R6.x.x/ R7.x.x) (cont’d) 4.) Customer will receive a request of Service side, click Yes to accept. See Figure 7-30 on page 7-26 Figure 7-30 Accept requirement of Service 5.) Service can do Remote Diagnostics successfully.
Page 175
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-7-3 Insite II Configuration (For LOGIQ e R6.x.x/R7.x.x and LOGIQ e Vet R6.x.x/ R7.x.x) (cont’d) 6.) Press power switch, select Full Maintenance Reboot to reboot system, see Figure 7-32 on page 7- Figure 7-32 Full Maintenance Reboot window 7.) In remote server side, select the device type in DEVICE TYPE.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-7-3 Insite II Configuration (For LOGIQ e R6.x.x/R7.x.x and LOGIQ e Vet R6.x.x/R7.x.x) (cont’d) 8.) Select Update Software in the category of SERVICE. Figure 7-34 Update software 9.) The page of software package shows up, select the correct software package in the list.
Page 177
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-7-3 Insite II Configuration (For LOGIQ e R6.x.x/R7.x.x and LOGIQ e Vet R6.x.x/R7.x.x) (cont’d) 10.)Click “+” icon on the schedule page to set up a schedule. Fill in the name and time, then click Add. Figure 7-36 New schedule page 11.)Then the active schedule is shown on the SCHEDULE page.
Page 178
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-7-3 Insite II Configuration (For LOGIQ e R6.x.x/R7.x.x and LOGIQ e Vet R6.x.x/R7.x.x) (cont’d) 12.)In device side, if the device is connected to network, a mail icon will show in the status bar. Figure 7-38 Status bar 13.)Press Power button.
Page 179
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-7-3 Insite II Configuration (For LOGIQ e R6.x.x/R7.x.x and LOGIQ e Vet R6.x.x/R7.x.x) (cont’d) NOTE: Make sure the system administrator is logged on as ADM, or the system will indicate such message.
Page 180
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-7-3 Insite II Configuration (For LOGIQ e R6.x.x/R7.x.x and LOGIQ e Vet R6.x.x/R7.x.x) (cont’d) 15.)When the download is complete, it will show the window below. Figure 7-43 Download Complete window 16.)Restart the system, and the system CD will be loaded as shown in Figure 7-44 on page 7-32 Figure 7-44 System CD loading...
Page 181
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 7-7-3 Insite II Configuration (For LOGIQ e R6.x.x/R7.x.x and LOGIQ e Vet R6.x.x/R7.x.x) (cont’d) 17.)After the system is on, login as ADM. Press Power button, and the system-exit window displays. Select Verify.
Page 182
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual This page was intentionally left blank. 7-34 Section 7-7 - Network and Insite II Configuration...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter describes replacement procedures for the following modules and subsystems. Table 8-1 Contents in Chapter 8 Section Description Page Number...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 8-2-2 Handle Assy (FRU No. 313) Purpose: This is a description on how to remove and replace the Handle Assy (FRU No.313). 8-2-2-1 Tools • Plier 8-2-2-2 Needed Manpower •...
Page 185
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 8-2-2 Handle Assy (FRU No. 313) (cont’d) Figure 8-1 Handle Assy Disassembly Chapter 8 Replacement Procedures...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 8-3 Loading the System Software NOTE: System Software may also be referred to as the Base Image or Ghost. While the software install procedure is designed to preserved data, you should save WARNING WARNING any patient data, images, system setups and customer presets to CD, DVD, USB Flash...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Loading Base Image Software with DVD 8-3-1 NOTE: While it is believed to be unnecessary, It would not hurt to disconnect the system from the network and remove all transducers. NOTE: Please ensure AC adapter is connected during system upgrade! 1.) Insert the disk labeled “System &...
Page 188
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 8-3-1 Loading Base Image Software with DVD (cont’d) 6.) Select one of the options for loading the system. Select choice [a] to load the complete disk. Refer to Figure 8-4 on page 8-6 If you select [a], ALL existing software and data will be erased.
Page 189
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 8-3-1 Loading Base Image Software with DVD (cont’d) 8.) System DVD will be loaded as shown in Figure 8-6 on page 8-7 Figure 8-6 System DVD loading If the screen displays as show in Figure 8-7 on page 8-7 after input “Yes”, wait about 1 minute until...
Page 190
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 8-3-1 Loading Base Image Software with DVD (cont’d) 9.) After finish updating system, remove the DVD-RW and press any key to shut down the system. Refer to Figure 8-8 on page 8-8 Figure 8-8 System upgrade complete 10.)Press any key to reboot system.
Page 191
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 8-3-1 Loading Base Image Software with DVD (cont’d) NOTE: For LOGIQ e R5.x.x, R6.x.x, please select a package to run the application. For Emergency&Critical For Nerve&Needle For General Imaging Figure 8-11 Package selection message NOTE: For LOGIQ e R5.x.x, you need to select a package to run the application.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 8-3-2 Loading the System Software with USB Memory Stick NOTE: Before starting this procedure, remove all probes and peripherals and remove them from the docking cart. NOTE: Please ensure AC adapter is connected during system upgrade! 1.) Properly turn off the scanner by momentarily pressing the Power On/Off Switch.
Page 193
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 8-3-2 Loading the System Software with USB Memory Stick (cont’d) 6.) The system software load instruction window displays. Select one of the options for loading the system. Select choice [a] to load the complete disk. Refer to Figure 8-14 on page 8-11 If you select [a], ALL existing software and data will be erased.
Page 194
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 8-3-2 Loading the System Software with USB Memory Stick (cont’d) 7.) Input “Yes” or “No” and press Enter key to continue. Figure 8-15 Confirmation on loading the system 8-12 Section 8-3 - Loading the System Software...
Page 195
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 8-3-2 Loading the System Software with USB Memory Stick (cont’d) 8.) System USB memory stick will be loaded as shown in Figure 8-16 on page 8-13 . Wait for the software installation to complete.
Page 196
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 8-3-2 Loading the System Software with USB Memory Stick (cont’d) 9.) After finish updating system, remove the USB memory stick and press any key to shut down the system.Refer to Figure 8-18 on page 8-14 NOTE:...
Page 197
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 8-3-2 Loading the System Software with USB Memory Stick (cont’d) 10.)Press any key to reboot system. When system first time boot up after upgrading complete, wait until the below screen displays. Select the package and select “Next“ to run the application; or to abort, select “shutdown“...
Page 198
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 8-3-2 Loading the System Software with USB Memory Stick (cont’d) 11.)When select “Next“ in step 10, the system will instruct to reboot the system to active the package as shown in Figure 8-20 on page 8-16 .
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 8-3-3 Reload the Correct Preset Region NOTE: After the system software loading completion, please reload the correct preset region. 1.) Reboot the system. 2.) Please select Utility-System-About and then select Additional About Information Figure 8-21 Additional About Information NOTE: After you have selected Utility-System-About and selected Additional About Information,...
Page 200
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 8-3-3Reload the Correct Preset Region (cont’d) 4.) The system will pop up a message about the selected region preset. Check if the selected region preset is correct, press OK. Figure 8-23 Preset region information 5.) Press OK to reboot the system.
Page 201
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 8-3-3Reload the Correct Preset Region (cont’d) 6.) After the system reboots, select Utility-System-About and select Additional About Information to check if the correct preset region is displayed. Figure 8-25 Preset region Chapter 8 Replacement Procedures 8-19...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 8-3-4 Option Strings Check NOTE: After the system software loading completion, please check the option strings to ensure that the options are activated and working. 1.) Reboot the system. 2.) Select Utility-Admin-System Admin 3.) Ensure that all the installed Option keys are displayed and the status of Options are valid.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 8-3-5 Probe Recognition Check NOTE: After the system software loading completion, please check to ensure that the system can recognize the probes. Plug in the probe. In scanning mode, the probe information is displayed on the right top location of the screen.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 8-3-7 Reinstall DICOM Devices Reinstall any DICOM devices used by the customers and check to ensure these DICOM devices work properly. The instruction about installing DICOM devices is not incorporated in this manual. To access the instruction about installing DICOM devices please refer to another manual Basic User Manual.
Page 205
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 8-4Checks after FRU replacement (Debrief Guidelines) Perform required Functional tests based upon the FRU being replaced. Table 8-2 Functional Tests Matrix of LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Section Description Functional Tests...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual This page was intentionally left blank. 8-24 Section 8-4 - Checks after FRU replacement (Debrief Guidelines)
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives an overview of Spare Parts available for the LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e . Table 9-1 Contents in Chapter 9 Section...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 9-3Renewal Parts Lists NOTE: The part replacement is shown by the “Replaced By” column in the table. If the part is replaced by a new version, the new version is shown in the “Replaced By” column in the table. The item number for the new version will have a letter in the alphabetical order after the Arabic numerals.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 9-5 LCD Assy Figure 9-2 LCD Assy Section 9-5 - LCD Assy...
Page 211
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 9-5 LCD Assy (cont’d) Table 9-3 LCD Assy Part Replaced Item Number Part Name LCD Assy for LOGIQ e R7.x.x / LOGIQ e Vet R6.0.5 or above/ 5419295 5419295-2 LOGIQ e Vet R7.x.x...
Page 212
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 9-3 LCD Assy Part Replaced Item Number Part Name Inverter Assy for LOGIQ e R4.x.x / Vivid e R4.x.x / LOGIQ i R4.x.x / 5147731 5147731-2 LOGIQ i R5.x.x Inverter Assy LOGIQ e R5.0.x / LOGIQ e R5.2.x/LOGIQ e R6.x.x / 103A...
Page 213
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 9-6Keyboard Assy Figure 9-3 Keyboard Assy Chapter 9 Renewal Parts...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 9-6 Keyboard Assy (cont’d) Figure 9-4 Keyboard Assy (cont’d) Section 9-6 - Keyboard Assy...
Page 215
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 9-4 Keyboard Assy Part Replaced Item Number Part Name Keyboard Assy for LOGIQ e R7.x.x / LOGIQ e Vet R6.0.5 or above/ 5419297 LOGIQ e Vet R7.x.x 5458515, 5439794, 5155004,...
Page 216
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 9-4 Keyboard Assy Part Replaced Item Number Part Name Keyboard cover Assy for LOGIQ e R7.x.x / LOGIQ e Vet R6.0.5 or 5419296 above/ LOGIQ e Vet R7.x.x Keyboard cover Assy for LOGIQ e R5.0.x / LOGIQ e R5.2.x /LOGIQ 5199756 5439794...
Page 217
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 9-7Bottom Assy 302A Figure 9-5 Bottom Assy Chapter 9 Renewal Parts 9-11...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 9-7 Bottom Assy (cont’d) Figure 9-6 Bottom Assy (cont’d) 9-12 Section 9-7 - Bottom Assy...
Page 219
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 9-7 Bottom Assy (cont’d) 316A Figure 9-7 Bottom Assy (cont’d) Chapter 9 Renewal Parts 9-13...
Page 220
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 9-7 Bottom Assy (cont’d) Table 9-5 Bottom Assy Part Replaced Item Number Part Name Bottom Cover Assy for LOGIQ e R4.x.x / LOGIQ e R5.0.x / 5148765 LOGIQ e R5.2.x /LOGIQ e R6.x.x/ LOGIQ e Vet R5.x.x/ R6.x.x...
Page 221
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 9-5 Bottom Assy Part Replaced Item Number Part Name RX64 Board for LOGIQ e R4.x.x LOGIQ e R5.x.x/ LOGIQ e 303A 2404906-2 R6.x.x/Vivid e R4.x.x/ Vivid e R5.x.x/ Vivid e R6.x.x /LOGIQ i R4.x.x/ LOGIQ i R5.x.x/ LOGIQ e Vet R6.x.x 2404906-3 RX64 Board for LOGIQ e R7.x.x/ LOGIQ e Vet R7.x.x...
Page 222
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 9-5 Bottom Assy Part Replaced Item Number Part Name U7500 CPU kits only for LOGIQ e R6.x.x / Vivid e R6.x.x/ 5392210-3 LOGIQ e Vet 6.x.x 307A U7500 CPU kits for LOGIQ e 7.x.x (Update BIOS)/ LOGIQ e 5392210-6...
Page 223
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 9-5 Bottom Assy Part Replaced Item Number Part Name SATA HDD with Blue Front Shell and Grub Patch Installation Kit for LOGIQ e R4.0.x / LOGIQ e R5.x.x /LOGIQ e R7.0.4 or below / Vivid e R4.0.x / Vivid e R5.0.x / Vivid e R6.0.0 / Vivid 5477476 e R6.0.1 /Vivid e R6.0.2 /LOGIQ e Vet R5.0.x / LOGIQ e Vet...
Page 224
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 9-8 Options of spare parts for LOGIQ e Vet Systems for USA NOTE: The FRUs of LOGIQ e Vet systems for USA need to be distinguished. •...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 9-8 Options of spare parts for LOGIQ e Vet Systems for USA (cont’d) • In case of LOGIQ e Vet with blue & gray color scheme: Figure 9-9 Blue &...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 9-10Isolation Cart Components Figure 9-10 Isolation Cart Components Table 9-8 Isolation Cart Components Part Replaced Item Number Part Name 5183729 Wheels 5180439-2 Isolation transformer 110V 5180376 Isolation transformer 200V~240V 5180376-2 502A...
Page 228
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 9-8 Isolation Cart Components Part Replaced Item Number Part Name 5182671 USA class cable kits 5182937 European class cable kits 5182252 Chinese class cable kits 5182323 Japanese class cable kits 5182095 Australia/New Zealand class cable kits...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 9-11 Isolation Cart Enhanced Version Components Figure 9-11 Isolation Cart Components Chapter 9 Renewal Parts 9-23...
Page 230
515A 5461076 Isolation Cart Rear Handle (Onyx Black) 5391606 Spring Cable Assy 5394036 Locate block and Screw cap (GE Pearl Metallic) 5426645 517A 5426645 Locate block and Screw cap (GE N9) 5394032 Probe and gel holder kit (Dark Steel Blue)
Page 231
AC Power Cord China 5177123-2 AC Power Cord Europe 5177146-2 AC Power Cord USA 5393025 Isolation Cart Drawer Kits (GE Pearl Metallic) 5426647 Isolation Cart Drawer Kits (GE N9) 5393026 Isolation Cart Basket Kits (GE Pearl Metallic) 5426648 Isolation Cart Basket Kits (GE N9)
Page 232
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 9-12 Accessories and Kits Table 9-10 Accessories and Kits Replaced Item Part Number Part Name Battery Pack for LOGIQ e R4.x.x / LOGIQ e R5.0.x / LOGIQ e R5.2.x /LOGIQ e R6.x.x/ LOGIQ e R7.x.x / LOGIQ e Vet / 5120410 5120410-2...
Page 233
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 9-10 Accessories and Kits Replaced Item Part Number Part Name 604D 5151234-5 PLEXTOR PX-L890UE DVDRW (EUR) 5449275-2 5182891-2 DVD-RW (JPN) LITEON DVD-RW 5449275-2 605A 5182891-3 DVD-RW (JPN) LITEON DVD-RW 5449275-2 605B 5182891-4...
Page 234
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 9-10 Accessories and Kits Replaced Item Part Number Part Name 5175689 HP470 Printer Kits (HP470+PIT) -US 5426597 617A 5426597 HP Officejet 100 Printer kit - USA 5175988 HP470 Printer Kits (HP470+PIT) -EU 5426595...
Page 235
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 9-10 Accessories and Kits Replaced Item Part Number Part Name 5149641 ECG Module with Chinese Label (China) ECG Module with SKD Label, only applicable for Muslim 5195563 countries ECG Module with SKD Label for Service Parts (other...
Page 236
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 9-10 Accessories and Kits Replaced Item Part Number Part Name 5196495 Clamp Filter for LOGIQ i R5.1.x 5268864 Peripheral Driver Patch CD Installation kit 5453562-5 642A 5455722 Peripheral Driver Patch CD Installation kit 5453562-5...
Page 237
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 9-10 Accessories and Kits Replaced Item Part Number Part Name 5322236 System and Application software DVD for LOGIQ e (R 5.2.0) 5322457-4 System and Application software DVD for LOGIQ e (R 5.2.1) 5322457-6 System and Application software DVD for LOGIQ e (R 5.2.2) 5322457-7...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 9-10 Accessories and Kits Replaced Item Part Number Part Name System and Application software DVD for LOGIQ e Vet (R 5248829 5.0.0) System and Application software DVD for LOGIQ e Vet (R 5248829-2 5.0.1) System and Application software DVD for LOGIQ e Vet (R...
Page 239
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 9-10 Accessories and Kits Replaced Item Part Number Part Name 5248828 System and Application software DVD for Vivid e (R 5.0.0) 5248828-2 System and Application software DVD for Vivid e (R 5.0.1) 5248828-3 System and Application software DVD for Vivid e (R 5.0.1) 5248828-4...
Page 240
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 9-10 Accessories and Kits Replaced Item Part Number Part Name 5240781-3 Transformer Service Kit - 110V 668A 5240780-2 Transformer Service Kit - 220V 5240780-3 5240780-3 Transformer Service Kit - 220V 5135447-R 669A 5135447...
Page 241
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 9-13Manuals Table 9-11 MANUALS for LOGIQ e R4.x.x Item Part Name Part Number Description Quantity LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Service 5370626 -100 Service Manual Manual System User Manuals 5118586-100...
Page 242
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 9-12 MANUALS for LOGIQ e R5.0.x/LOGIQ e R5.2.x/LOGIQ e R6.x.x/R7.x.x/LOGIQ e Vet Item Part Name Part Number Description Quantity 5199656 -127 Basic User Manual (for R5.0.x) LOGIQ e User Manual, Portuguese 5199656 -140 Basic User Manual (for R5.0.x)
Page 243
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 9-12 MANUALS for LOGIQ e R5.0.x/LOGIQ e R5.2.x/LOGIQ e R6.x.x/R7.x.x/LOGIQ e Vet Item Part Name Part Number Description Quantity LOGIQ e R5.0.x / LOGIQ e Vet Quick 5212024 -106 Quick Start Guide (for R5.0.x) Start Guide, Spanish...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 9-13 MANUALS for Vivid e R4.x.x Item Part Name Part Number Description Quantity 5175726 -111 Quick Start Guide Vivid e R4.x.x Quick Start Guide, Italian 5175726 -127 Quick Start Guide Vivid e R4.x.x Quick Start Guide, Portuguese 5175726 -140...
Page 245
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 9-15 MANUALS for LOGIQ i R4.x.x / LOGIQ i R5.x.x Item Part Name Part Number Description Quantity System User Manuals 5179205-100 Basic User Manual LOGIQ i R4.x.x User Manual, English 5179205-101 Basic User Manual LOGIQ i R4.x.x User Manual, French...
Page 246
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 9-14Probe Table 9-16 Probes for LOGIQ e, Vivid e and LOGIQ i Part Replaced Part Item Name Description Number Probe ± 4C-RS 5131629 (Center Frequency: 3.1 20% MHz) Probe E8C-...
Page 247
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Table 9-16 Probes for LOGIQ e, Vivid e and LOGIQ i Part Replaced Part Item Name Description Number Probe ± 9L-RS 5213143 (Center Frequency: 5.25 5499511 20% MHz) Probe ±...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 9-14 Probe (cont’d) Table 9-17 Probes for LOGIQ e Vet Part Item Part Name Number Description ± 4C-RS Vet 5198378 Probe (Center Frequency: 3.1 20% MHz) ±...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Chapter 10 Care & Maintenance Section 10-1 Overview 10-1-1 Periodic Maintenance Inspections It has been determined by engineering that your LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e system does not have any high wear components that fail with use, therefore no Periodic Maintenance Inspections are mandatory.
Testing results, corrective action and the effects of corrective action must be documented and maintained on the site. Your GE service representative can help you with establishing, performing and maintaining records for a quality assurance program. Please contact us for coverage information and/or price for service.
Your GE Service Representative has an in-depth knowlegde of your LOGIQ e/LOGIQ e Vet/LOGIQ i/ Vivid e ultrasound scanning system and can best provide competent, efficient service. Please contact us for coverage information and/or price for service.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 10-4 Tools Required 10-4-1 Special Tools, Supplies and Equipment 10-4-1-1 Specific Requirements for Care & Maintenance Table 10-3 Overview of Requirements for Care & Maintenance Tool Part Number Comments Digital Volt Meter (DVM)
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 10-5 System Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary. Table 10-4 System Checks Step Item Description...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 10-5-2 Functional Checks (See Also Chapter 4) The functional checks take about 60 minutes to perform. Refer to the system user documentation whenever necessary. 10-5-2-1 System Checks Table 10-5 System Functional Checks ÷...
Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. Refer to the User Manual for a list of approved peripherals/options. Table 10-6 GE Approved Peripheral/Hardware Option Functional Checks Step Item Description B/W Printer Verify hardcopy output of the B/W video page printer.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 10-5-5 Physical Inspection Table 10-9 Physical Checks Step Item Description Verify that all system labeling is present and in readable condition. Refer to User Manual,..for Labeling details.
The system user manuals and various probe handling cards provide a complete description of probe care, maintenance, cleaning and disinfection. Ensure that you are completely familiar with the proper care of GE probes. Ultrasound probes can be easily damaged by improper handling. See the User Manual and probe care cards for more details.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 10-5-8 Battery Performance Maintenance Battery replacement every three years is recommended. It is recommended to do battery performance maintenance one time per year. Please follow the flow chart below to carry out battery performance maintenance. Battery discharge >45 minutes? Battery charge >...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 10-6 Electrical Safety Tests 10-6-1 Safety Test Overview The electrical safety tests in this section are based on and conform to IEC 60601-1 Medical Equipment Safety Standards. They are intended for the electrical safety evaluation of cord-connected, electrically operated, patient care equipment.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 10-6-2 GEMS Leakage Current Limits The following limits are summarized for IEC 60601-1 Medical Equipment Safety Standards. These limits are GEMS standards and in some cases are lower than the above standards listed. Table 10-11 Chassis Leakage Current Limits—Accessible Metal Surfaces Country Normal Condition...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 10-6-3 Outlet Test - Wiring Arrangement Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated. Any problems found should be reported to the hospital immediately and the receptacle should not be used.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 10-6-4 Chassis Leakage Current Test 10-6-4-1 Definition This test measures the current that would flow in a grounded person who touched accessible metal parts of the bedside station if the ground wire should break. The test verifies the isolation of the power line from the chassis.
Page 263
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 10-6-4-3 Data Sheet for enclosure Source Leakage Current The test passes when all readings measure less than the value shown in Table 10-11. Record all data on the PM Inspection Certificate. Table 10-14 Typical Data Sheet for enclosure Source Leakage Current Test 2 Tester Neutral...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 10-6-5 Probe Leakage Current Test 10-6-5-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface. 10-6-5-2 Generic Procedure Measurements should be made with the ground open and closed, with power line polarity normal and...
Page 265
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual 10-6-5-4 Data Sheet for Transducer Source Leakage Current The test passes when all readings measure less than the values shown in Table 10-11. Record all data on the PM Inspection Certificate. Equipment damage possibility.
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Section 10-7 When There's Too Much Leakage Current... AC/DC FAILS Check any broken of the AC/DC adapter and its cable. Replace a new one if any portion defective. ENCLOSURE FAILS Check any broken of the enclosure.
Page 267
B-Mode Function Console Doppler Modes Function CF-Mode Function External I/O M-Mode Function Cables and Connectors Applicable Software Options GE Approved Peripherals (DVD-RW, Printer) Applicable Hardware Options Labeling (see User Manual for Labeling) Control Panel Measurement Accuracy GE Approved Peripherals COMMENTS: ______________________________________________________________________________ ______________________________________________________________________________ Chapter 10 Care &...
Page 268
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual ______________________________________________________________________________ ______________________________________________________________________________ ELECTRICAL SAFETY Max Value Value Electrical Test Performed Allowed Measured Comments Outlet (correct ground &wiring config.) Type BF Applied Part Leakage Current Limits- Probe enclosure Source Leakage Current - Chassis Leakage Current Limits Peripheral 1 Leakage Current Peripheral 2 Leakage Current...
Page 269
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual INDEX EMI, 1-17 ESD, 1-17 Abbreviations, 9-1 Exam study Archiving Images display location, 4-9 Move, 4-31 Backup Patient Database, 4-28 Preset Configurations, 4-28 Basic Measurements Focal zone Functional Checks, 4-25 display location, 4-9 Body pattern...
Page 270
GE H EALTHCARE Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual Measurement summary window Safety Considerations, 1-13 display location, 4-9 Screen Captures, 7-6 Mechanical Safety, 1-13 System Maintenance, 10-5 Media System Manufacturer, 1-19 Formatting, 4-28...
Need help?
Do you have a question about the LOGIQ e R4 Series and is the answer not in the manual?
Questions and answers