3D and 4D Mode
8.8.6.3 SonoAVC™antral2.0
Note
Note
Volume Analysis Menu: SonoAVC™
ROI shape
Start SonoAVC™
Ref. Image
8-42
yes: continue with next step
Add to Report to save the measurements to the current exam.
7.
Select
Follicle measurement method:
1.
Scan and freeze a follicle volume data set.
2.
Select Volume Analysis and select SonoAVC™
3.
Adjust the ROI.
auto and adjust the ROI shape if desired.
4.
Select
5.
Start the measurement by selecting either
6.
The rendered follicles and the result list are displayed on screen.
7.
Edit the results if necessary.
Add to Report to save the measurements to the current exam.
8.
Select
SonoAVC™antral
2.0
is an option.
If a 4D Volume cine is present, the system will automatically switch to 3D Static when
SonoAVC™antral is pressed.
"Antral" means "antral follicle count". SonoAVC™
count antral follicles within a ROI box in a 3D volume data set.
antral
2.0
Figure 8-28 Volume Analysis Menu: SonoAVC™antral
Select the desired ROI shape:
angular : the box has the shape of a rectangle
rounded : the box has an elliptic shape with rounded corners that can be
adjusted by pressing (default position) and rotating (rounding the corners) the
rotary control. It is embedded in a rectangular box.
Left Ovary or Right Ovary to start SonoAVC™.
Select
Select the reference image to which all image dependent functions like parallel
shifts, rotations, etc. are applied.
auto
follicle .
Left Ovary or Right Ovary .
antral
2.0
2.0
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
enables to automatically detect and