Using M-Mode; M-Mode Menu - GE Voluson Swift Instructions For Use Manual

Table of Contents

Advertisement

Color
Flow Profiles
Info
Cine Menu
6.2.4 M-Mode

Using M-Mode

Hint
Info
Note
M-Mode Main Menu

M-Mode Menu

Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 Revision 4
Color on/off.
select between 3 different buttons
active at the same time. With pressing the button again it will be deactivated. Use
Flow Presets for optimizing vessel flow representation scanning.
Also see 'B-Mode' on page 6-4 and 'Button description' on page 3-5.
For more information see 'Cine Mode' on page 7-12.
M-Mode is intended to provide a display format and measurement capability that represents
tissue displacement (motion) occurring over time along a single vector.
M-Mode is used to determine patterns of motion for objects within the ultrasound beam. The
most common use is for viewing motion patterns of the heart.
1.
Touch 2D on the user interface to start B-Mode.
2.
Touch M on the user interface to start M-Mode.
3.
The
M Main menu appears .
4.
Place the cursor line over the region of interest.
5.
Press
2D/M run (right or left trackball button).
Freeze .
6.
Press
To change M Gain rotate the 2D-button. To change the Frequency use the touch panel.
Also see 'Button description' on page 3-5. For the option AMM (Anatomical M-Mode) please
see 'Anatomical M-Mode (AMM)' on page 6-12.
The availability of some functions or features depends on the probe and ultrasound system
used.
Display Format
Different options for display arrangement.
AMM
Opens the AMM mode.
Dyn. Contr.
Dynamic Contrast controls how echo intensities are
converted to shades of gray, thereby increasing the
adjustable range of contrast.
Harm. Frequ.
Multi Frequency mode lets you downshift to the probe's
next lower frequency or shift up to a higher frequency.
Cine #
Select the desired Cine #.
Display Format
Different options for display arrangement.
AMM
Opens the AMM mode.
Dyn. Contr.
Dynamic Contrast controls how echo intensities are
converted to shades of gray, thereby increasing the
adjustable range of contrast.
UtA , UmA and MCA . Only one button can be
2D Mode
6-11