The Voluson™ SWIFT / Voluson SWIFT+ is intended for use in electromagnetic environment specified below. The
customer or the user of the Voluson™ SWIFT / Voluson SWIFT+ should assure that it is used in such an environment.
Immunity Type
IEC/EN 61000-4-6
Conducted RF
IEC/EN 61000-4-3
Radiated RF EM fields
Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 Revision 4
Guidance and manufacturer´s declaration - electromagnetic immunity
Equipment Capability
3 Vms at 0.15 MHz- 80
MHz,
80% AM at 1kHz
6 Vms in ISM bans between
0.15 MHz - 80 MHz
80% AM at 1kHz
3V/m
80MHz - 2.7GHz
80 % AM at 1 kHz
Regulatory Acceptable Level
Edition 3
3 Vrms at 0.15 MHz - 80
MHz,
80 % AM at 1 kHz
Edition 4
3 Vrms at 0.15 MHz - 80
MHz
6 Vrms in ISM bands
between 0.15 MHz - 80 MHz
80% AM at 1kHz
Edition 3
3V/m, 80 MHz – 2.5 GHz,
80% AM at 1 kHz
Edition 4
3V/m
80MHz - 2.7GHz
80 % AM at 1 kHz
Safety
Electromagnetic
environment - guidance
Portable and mobile RF
communications equipment
should be used no closer to
any part of this system ,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance.
where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in metres (m). Field
strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey, a) should be less
than the compliance level in
each frequency range. b)
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
2-23
Need help?
Do you have a question about the Voluson Swift and is the answer not in the manual?