IT (Intracranial Translucency)

10.3.1.3 SonoBiometry

10.3.1.4 Facial Angle Measurement

Note

10.3.1.5 Sono FHR

Voluson™ SWIFT / Voluson SWIFT+ Instructions For Use
5831612-100 Revision 4
SonoIT (Sonography based Intracranial Translucency) is a system supported
measurement for Intracranial Translucency. Starting from the routinely used
midsagittal view of the fetal face, obtained for assessment of the Nuchal
Translucency and nasal bone, the ultrasound system uses a semi-automated
mode to measure the anterior-posterior diameter of the fourth ventricle
recognizable as intracranial translucency.
The workflow is identical with SonoNT.
Biometry is an alternative to the common fetal biometry measurements. It provides
Sono
system suggested measurements for BPD, HC, AC, FL,HL, Cerebellum, CM or VP which need
to be confirmed by the user or can be changed manually.
To use :
1.
Press Calc on the user interface.
2.
Select the measurement package
3.
Select the desired measurement item (
4.
The calculation process starts. The result is displayed on the monitor screen. If the result
is not correct, proceed with manual correction by pressing
trackball cursor.
5.
Press Set to accept the result and to finish the measurement.
Two facial angle measurements are available:
FMF Angle (Frontomaxillary Facial Angle): The FMF angle is measured between a line
along the upper surface of the palate and the upper corner of the anterior aspect of the
maxilla, extending to the external surface of the forehead, represented by the frontal
bones or an echogenic line under the skin below the metopic suture that remains open.
MMF Angle (Mandibulomaxillary Facial Angle): The MMF angle is generated by using
the same first part and the same apex as for the FMF angle. The second part, however,
is drawn downwards and positioned so that the inner aspect of the line flushes with the
upper anterior corner of the mandible.
The Facial Angle measurements are not included in the presets but need to be added
manually to a measurement group.
Figure 10-4 Monitor display: Fetal Facial Angle measurement
Measurements and Calculations
OB .
BPD , HC , AC , FL , HL , Cerebellum , CM , or VP ).