Program Subsystem - Chroma 63200A Series Operation & Programming Manual

High power dc electronic loads
Table of Contents

Advertisement

High Power Electronic Load 63200A Series Operation & Programming Manual
4.3.2.13

PROGRAM Subsystem

PROGram:DATA
Type:
Description:
Setting Syntax:
Setting Parameters:
Setting Example:
Query Syntax:
Return Parameters:
Query Example:
Return Example:
PROGram:DATA:LIST
Type:
Description:
Setting Syntax:
Setting Parameters:
4-54
Channel-Specific
Set the program parameters. (Note: All setting parameters in this
command can't use suffix.)
PROGram:DATA<space><Arg1>,<Arg2>,<Arg3>,<Arg4>,<Arg5>
Selects a program to be set:
Arg1: <NR1>, 1 ~ 10, Resolution = 1, Unit = None.
Set the type of program:
Arg2: <NRf>, LIST(0), STEP(1), Unit = None.
Set the chain parameter in program:
Arg3: <NR1>, 0 ~ 10, Resolution = 1, Unit = None.
Set the repeat count of program:
Arg4: <NR1>, 0 ~ 9,999, Resolution = 1, Unit = None.
Set number of sequence in program:
Arg5: <NR1>, 0 ~ 100, Resolution = 1, Unit = None.
PROG:DATA 1,STEP,2,0,5
PROGram:DATA?<space><NR1>[<space><MAX | MIN>]
<aard>
PROG:DATA? 1
PROG:DATA? 1 MAX
PROG:DATA? 1 MIN
1,LIST,3,1,5
Channel-Specific
Set the list parameters in program. (Note: All setting parameters in
this command can't use suffix.)
PROGram:DATA:LIST<space><Arg1>,<Arg2>,<Arg3>,<Arg4>,
<Arg5>,<Arg6>,<Arg7>,<Arg8>,<Arg9>,<Arg10>,<Arg11>,
<Arg12>,<Arg13>,<Arg14>,<Arg15>,<Arg16>
Selects a program to be set:
Arg1: <NR1>, 1 ~ 10, Resolution = 1, Unit = None.
Selects a sequence to be set:
Arg2: <NR1>, 1 ~ N, Resolution = 1, Unit = None.
Set the trigger mode of sequence:
Arg3: <NRf>, SKIP(0), AUTO(1), MANUAL(2), EXTERNAL(3),
Unit = None.
Set the run mode of sequence:
Arg4: <NRf>, CC(0), CR(1), CV(2), CP(3), Unit = None.
Set the mode's range of sequence:
Arg5: <NRf>, LOW(0), MIDDLE(1), HIGH(2), Unit = None.
Set the load value according to run mode in sequence:
Arg6: <NRf>, Refer to respective specification for valid value
range.
Set the falling of slew rate in sequence:
Arg7: <NRf>, Refer to respective specification for valid value
range.
Set the rising of slew rate in sequence:
Arg8: <NRf>, Refer to respective specification for valid value