High Power Electronic Load 63200A Series Operation & Programming Manual
POWer:STATic:FALL
Type:
Description:
Setting Syntax:
Setting Parameters:Refer to respective specification for valid value range.
Setting Example:
Query Syntax:
Return Parameters:<NR2>, [Unit = A/μs]
Query Example:
POWer:STATic:VRNG
Type:
Description:
Setting Syntax:
Setting Parameters:<CRD | NR1>, LOW | L | 0, MIDDLE | M | 1, HIGH | H | 2
Setting Example:
Query Syntax:
Return Parameters:<CRD>, LOW(0), MIDDLE(1), HIGH(2)
Query Example:
4.3.2.8

ADVANCE Subsystem

[ADVance:]BATTery:MODE
Type:
Description:
Setting Syntax:
Setting Parameters:<CRD | NR1>, CC(0), CR(1), CP(2)
Setting Example:
Query Syntax:
Return Parameters:<CRD>, CC(0), CR(1), CP(2)
Query Example:
[ADVance:]BATTery:VALue
Type:
Description:
Setting Syntax:
Setting Parameters:Refer to respective specification for valid value range.
Setting Example:
4-26
Channel-Specific
Set the falling slew rate of current for constant power mode.
POWer:STATic:FALL<space><NRf+>[suffix]
POW:STAT:FALL 2.5
POW:STAT:FALL 1A/μs
POW:STAT:FALL MAX
POW:STAT:FALL MIN
POWer:STATic:FALL?[<space><MAX | MIN>]
POW:STAT:FALL?
POW:STAT:FALL? MAX
POW:STAT:FALL? MIN
Channel-Specific
Set the voltage measurement range in constant power mode.
POWer:STATic:VRNG<space><CRD | NR1>
POW:STAT:VRNG HIGH
POW:STAT:VRNG M
POW:STAT:VRNG 0
POWer: STATic: VRNG?
POW:STAT:VRNG?
Channel-Specific
Set run mode in battery discharge mode.
[ADVance:]BATTery:MODE<space><CRD | NR1>
BATT:MODE CC
BATT:MODE 2
[ADVance:]BATTery:MODE?
BATT:MODE?
Channel-Specific
Set load value according to the run mode in battery discharge mode.
[ADVance:]BATTery:VALue<space><NRf+>[suffix]
When BATT:MODE set to CC mode, then
BATT:VAL 0.5
BATT:VAL 500mA
BATT:VAL MAX
Set falling slew rate to 2.5A/μs.
Set falling slew rate to 1A/μs.
Set falling slew rate to the
maximum value.
Set falling slew rate to the
minimum value.
Set voltage range to High.
Set voltage range to Middle.
Set voltage range to Low.
[Unit = None]
Set run mode = CC
Set run mode = CP
[Unit = None]
Set current = 0.5A
Set current = 0.5A
Set current = maximum value.