Chroma 63200A Series Operation & Programming Manual page 63

High power dc electronic loads
Table of Contents

Advertisement

High Power Electronic Load 63200A Series Operation & Programming Manual
To enter into EXTW mode, press
MODE: Able to set to CC, CR or CV mode.
CC mode
Ext
Iset
10
CR mode
10
Rset
Ext
CV mode
Ext
Vset
10
I
_
Limit
Ranges (Low, Middle, High)
It can be programmed in any of the three ranges, low range, middle range and high range.
The low range provides better resolution. If any value is over the maximum of low range,
you must select the middle range. When any value is over the maximum of middle range,
you must select the high range. To change the range, press the
the VFD range indicator is active at you want to select. The mode change will affect the Load,
so will the change of range. Both of them will cause the input to go through an off state.
The new setting will change the input immediately at a rate determined by the slew rate
setting.
3-32
ADVA
_
I
I
F
.
S
.
V
V
R
F
.
S
.(min)
_
V
_
V
V
F
.
S
.
V
Ext
_
I
I
F
.
S
.
10
V
and select EXTW.
RANGE
key few times until