Chroma 63200A Series Operation & Programming Manual page 79

High power dc electronic loads
Table of Contents

Advertisement

High Power Electronic Load 63200A Series Operation & Programming Manual
Query Syntax:
Return Parameters:<NR2>, [Unit = A/μs]
Query Example:
CURRent:STATic:FALL
Type:
Description:
Setting Syntax:
Setting Parameters:Refer to respective specification for valid value range.
Setting Example:
Query Syntax:
Return Parameters:<NR2>, [Unit = A/μs]
Query Example:
CURRent:STATic:VRNG
Type:
Description:
Setting Syntax:
Setting Parameters:<CRD | NR1>, LOW | L | 0, MIDDLE | M | 1, HIGH | H | 2
Setting Example:
Query Syntax:
Return Parameters:<CRD>, LOW(0), MIDDLE(1), HIGH(2)
Query Example:
CURRent:DYNamic:L1
Type:
Description:
Setting Syntax:
Setting Parameters:Refer to respective specification for valid value range.
Setting Example:
Query Syntax:
Return Parameters:<NR2>,[Unit = Ampere]
4-16
CURR:STAT:RISE MIN
CURRent:STATic:RISE?[<space><MAX | MIN>]
CURR:STAT:RISE?
CURR:STAT:RISE? MAX
CURR:STAT:RISE? MIN
Channel-Specific
Set the falling slew rate of current for constant current static mode.
CURRent:STATic:FALL<space><NRf+>[suffix]
CURR:STAT:FALL 2.5
CURR:STAT:FALL 1A/μs
CURR:STAT:FALL MAX
CURR:STAT:FALL MIN
CURRent:STATic:FALL?[<space><MAX | MIN>]
CURR:STAT:FALL?
CURR:STAT:FALL? MAX
CURR:STAT:FALL? MIN
Channel-Specific
Set the voltage measurement range in CC mode.
CURRent STATic:VRNG<space><CRD | NR1>
CURR:STAT:VRNG HIGH
CURR:STAT:VRNG M
CURR:STAT:VRNG 0
CURRent: STATic: VRNG?
CURR:STAT:VRNG?
Channel-Specific
Set the load current during T1 period for constant current dynamic
mode.
CURRent:DYNamic:L1<space><NRf+>[suffix]
CURR:DYN:L1 20
CURR:DYN:L1 10A
CURR:DYN:L1 MAX
CURR:DYN:L1 MIN
CURRent:DYNamic:L1?[<space><MAX | MIN>]
Set rising slew rate to the
minimum value of static load.
Set falling slew rate to 2.5A/μs.
Set falling slew rate to 1A/μs.
Set falling slew rate to the
maximum value of static load.
Set falling slew rate to the
minimum value of static load.
Set voltage range to High.
Set voltage range to Middle.
Set voltage range to Low.
[Unit = None]
Set the dynamic load parameter
L1 = 20A.
Set the dynamic load parameter
L1 = 10A.
Set the dynamic load parameter
L1 = maximum value.
Set the dynamic load parameter
L1 = minimum value.