Voltage Subsystem - Chroma 63200A Series Operation & Programming Manual

High power dc electronic loads
Table of Contents

Advertisement

Setting Syntax:
Setting Parameters:Refer to respective specification for valid value range.
Setting Example:
Query Syntax:
Return Parameters:<NR2>, [Unit = A/μs]
Query Example:
RESistance:DYNamic:IRNG
Type:
Description:
Setting Syntax:
Setting Parameters:<CRD | NR1>, LOW | L | 0, MIDDLE | M | 1, HIGH | H | 2
Setting Example:
Query Syntax:
Return Parameters:<CRD>, LOW(0), MIDDLE(1), HIGH(2)
Query Example:
4.3.2.6

VOLTAGE Subsystem

VOLTage:STATic:L1
Type:
Description:
Setting Syntax:
Setting Parameters:Refer to respective specification for valid value range.
Setting Example:
Query Syntax:
Return Parameters:<NR2>, [Unit = Volt]
Query Example:
VOLTage:STATic:L2
Type:
Description:
Setting Syntax:
Setting Parameters:Refer to respective specification for valid value range.
Setting Example:
RESistance:DYNamic:FALL<space><NRf+>[suffix]
RES:DYN:FALL 2.5
RES:DYN:FALL 1A/μs
RES:DYN:FALL MAX
RES:DYN:FALL MIN
RESistance:DYNamic:FALL?[<space><MAX | MIN>]
RES:DYN:FALL?
RES:DYN:FALL? MAX
RES:DYN:FALL? MIN
Channel-Specific
Set the current measurement range in constant resistance dynamic
mode.
RESistance:DYNamic:IRNG<space><CRD | NR1>
RES:DYN:IRNG HIGH
RES:DYN:IRNG M
RES:DYN:IRNG 0
RESistance:DYNamic:IRNG?
RES:DYN:IRNG?
Channel-Specific
Set the static load voltage in constant voltage mode.
VOLTage:STATic:L1<space><NRf+>[suffix]
VOLT:STAT:L1 8
VOLT:STAT:L1 24V
VOLT:STAT:L1 MAX
VOLT:STAT:L1 MIN
VOLTage:STATic:L1?[<space><MAX | MIN>]
VOLT:STAT:L1?
VOLT:STAT:L1? MAX
VOLT:STAT:L1? MIN
Channel-Specific
Set the static load voltage in constant voltage mode.
VOLTage:STATic:L2<space><NRf+>[suffix]
VOLT:STAT:L2 8
Remote Operation
Set falling slew rate to 2.5A/μs.
Set falling slew rate to 1A/μs.
Set falling slew rate to the maximum
value of dynamic load.
Set falling slew rate to the minimum
value of dynamic load.
Set current range to High.
Set current range to Middle.
Set current range to Low.
[Unit = None]
Set voltage of load L1 as 8V.
Set voltage of load L1 as 24V.
Set voltage of load L1 as the
maximum value.
Set voltage of load L1 as the minimum
value.
Set voltage of load L2 as 8V.
4-23