Auto Mode; Setting A Program Sequence - Chroma 63200A Series Operation & Programming Manual

High power dc electronic loads
Table of Contents

Advertisement

Parameters:
V_SET: Set the voltage level.
R_SET: Set the resistance level.
This mode is suggested to be used in the UUT with CV output.
CAUTION
It is suggested to use a remote sense cable to measure the UUT output
voltage.
3.7.9

Auto Mode

In Auto mode, it has to program the constant voltage, constant resistance, constant current
and constant power, and then start the UUT for output. When the UUT voltage starts to
output, the Load will sink according to the programmed constant voltage in CV mode.
When the voltage rises it will automatically switch to CR mode and to the CC mode at last for
sinking. It will switch to CP mode for sinking if the UUT outputs high voltage abnormally.
To enter into Auto mode, press
Parameters:
V_SET: Set the voltage level.
R_SET: Set the resistance level.
I_SET: Set the current level.
P_SET: Set the power level.
Auto mode:
I
This mode is suggested to be used in the UUT with CV output.
CAUTION
It is suggested to use a remote sense cable to measure the UUT output
voltage.
3.7.10

Setting a Program Sequence

In PROG mode, the user can select the Electronic Load to do basic testing via the
programmed sequences. Also different program sequences can be linked for auto execution.
The function of program sequence is very powerful. The Electronic Load has 10 programs
that can set up 256 sequences maximum. For instance, when program 1 is set up with 5
sequences and program 2 is set up with 8 sequences, the rest programs from 3 to 10 can
set up the remaining 243 sequences. Or, it can set total 256 sequences in program 1.
Different sequences combination can be created through the program chain.
ADVA
and select AUTO and then press Enter.
CC
CR
CV
CP
Operation Overview