Advance Mode - Chroma 63200A Series Operation & Programming Manual

High power dc electronic loads
Table of Contents

Advertisement

REPEAT: Set the number of time to repeat (0=infinite loop).
I_RANGE: Set the current measurement range of Electronic Load. There are H, M and L for
selection.
Ranges (Low, Middle, High)
Resistance can be programmed in any of the three ranges, low range, middle range and
high range. The low range provides better resolution at low resistance setting. If any value
is over the maximum of low range, you must select the middle range. When any value is
over the maximum of middle range, you must select the high range. To change the range,
RANGE
press the
key few times until the VFD range indicator is active at you want to select.
The mode change will affect the Load, so will the change of range. Both of them will cause
the input to go through an off state. If the CRD mode of Load is active, the new setting will
change the input immediately at a rate determined by the slew rate setting.
UUT: 20V
Load1=5Ω, Load2=10Ω, SR / =0.2A/μs, SR\ =0.2A/μs, T1=10ms, T2=10ms, RT=0
Current
4A
2A
It is suggested to use a remote sense cable to measure the UUT output
CAUTION
voltage.
3.7

Advance Mode

The Electronic Load has useful advanced functions such as BATT and Sine Wave Dynamic
measurement, etc. Press
select the desired mode and press Enter.
10mS
10mS
10mS
T1
T2
Figure 3-7
ADVA
to enter into Advance page and use left/right arrow key to
10mS
T1
T2
Operation Overview
Load1
load2
Time
3-23