Synchronous Subsystem; Status Subsystem - Chroma 63200A Series Operation & Programming Manual

High power dc electronic loads
Table of Contents

Advertisement

Low byte ―――→ High byte
#3048<byte1><byte2><byte3><byte4><byte5><byte6><byte7><byte8>
Query Example:
Return Example:
4.3.2.14

SYNCHRONOUS Subsystem

SYNChronous:RUN
Type:
Description:
Setting Syntax:
Setting Parameters:<NRf>, OFF(0), ON(1)
Setting Example:
SYNChronous:TYPE
Type:
Description:
Setting Syntax:
Setting Parameters:<NRf>, NONE(0), MASTER(1), SLAVE(2)
Setting Example:
4.3.2.15

STATUS Subsystem

STATus:CHANnel:CONDition?
Type:
Description:
Query Syntax:
Return Parameters:
Bit Position
15
14
MAX_
Condition
LIM
Bit Weight
32768 16384 8192 4096 2048 1024
Query Example:
Return Example:
Sequence #1
PROG:SEQ:SPEC? 1,V
None
All Channels
Set all electronic loads to "ON" in sync. dynamic run.
SYNChronous:RUN<space><NRf>
SYNC: RUN ON
SYNC: RUN OFF Set the load to "OFF" on sync. parallel.
All Channels
Set the specified 63200A to master or slave for sync. dynamic run.
SYNChronous:TYPE<space><NRf>
SYNC:TYPE MASTER
SYNC:TYPE SLAVE
SYNC:TYPE NONE
Channel-Specific
Returns the real time channel status.
STATus:CHANnel:CONDition?
<NR1>
Bit Configuration of Channel Status Register
13
12
11
10
RMT
VCC FAN SYNC OTP OPP3 OPP2 OPP1 OCP3 OCP2 OCP1 REV OV2 OV1
_INH
STAT:CHAN:COND?
2048
Sequence #2
Set the load to "ON" on sync. parallel.
Set the 63200A to master for sync.
dynamic.
Set the 63200A to slave for sync.
dynamic.
Disables the 63200A to sync.
9
8
7
6
512
256
128
64
32
Return the status of the electronic load.
Remote Operation
、、、
5
4
3
2
16
8
4
1
0
2
1
4-59