High Power Electronic Load 63200A Series Operation & Programming Manual
632xxA-150-xxx
Checking Medium Voltage Range
A.
After tested the high voltage range, press
B.
The DC Source voltage outputs the voltage/current values listed in Table 6-6.
C.
Wait for 30 seconds after the DC Source outputted and to record the voltage measured
by DMM (V) and the Load.
632xxA-150-xxx
Checking Low Voltage Range
A.
After tested the medium voltage range, press
B.
The DC Source voltage outputs the voltage/current values listed in Table 6-7.
C.
Wait for 30 seconds after the DC Source outputted and to record the voltage measured
by DMM (V) and the Load.
632xxA-150-xxx
6-4
150V
15V
Table 6-6
DC Source
Model
Output Voltage
80V
8V
Table 6-7
DC Source
Model
Output Voltage
16V
1.6V
DMM (V)±0.045V
DMM (V)±0.0247V
RANGE
to M range.
Front Panel Display
Reading
DMM (V)±0.024V
DMM (V)±0.0132V
RANGE
to L range.
Front Panel Display
Reading
DMM (V)±0.0048V
DMM (V)±0.00264V