2.3.3 PRESET Parameter Setting

How to enter the PRESET menu
In [Standby Menu], press Function Key [PRESET] to enter the PRESET menu as shown
below:
1.ACV FREQUENCY:
2.SOFTWARE AGC :
3.WV AUTO-RANGE:
4.IR AUTO-RANGE:
5.GFI
50Hz/60Hz
(1) After entering the PRESET parameter setting menu, press Function Key [F3] to
move the highlight to the parameter item you desired to set.
(2) Press Function Key [F1] and [F2] to set the parameter data.
(3) PRESET parameter function table as shown below:
Item
ACV FREQUENCY
SOFTWARE AGC
WV AUTO-RANGE
IR AUTO-RANGE
GFI
FAIL RESTART
SCREEN

2.3.4 PROGRAM Parameter Setting

Test parameter setting
(1) In [Standby Menu], press Function Key [PROGRAM] to enter the PROGRAM
setting menu as the following:
STEP
1/1
VOLT :0.050kV
HIGH :0.500mA
TIME : 3.0s
1-10
Test steps setting
(1) Press [F3] to move the highlight and stay in the test item. When Function Key [F1]
shows "UP", it means press this key can switch to the next test step. When Function
60Hz
ON
OFF
ON
:
ON
RMT
LOCK OFST ERR
Range
Default
50 / 60 Hz 60Hz
OFF / ON
ON
OFF / ON
OFF It sets if to open the HiPot auto range function.
OFF / ON
ON
OFF / ON
ON
OFF / ON
OFF It sets if to restart directly when fails the test.
OFF/ON
ON
AC
LOW : OFF
ARC : OFF
RAMP : OFF
FALL : OFF
RMT
LOCK OFST ERR
UP
DOWN
NEXT
EXIT
Description
It sets the output voltage frequency at AC HiPot
test.
It sets if to open the software AGC function.
It sets if to open the insulation auto range
function.
It sets if to open the GFI function.
It sets if to display the test screen.
NEW
MORE..
NEXT
EXIT
Operation and Setting
2-9