Chroma 63200A Series Operation & Programming Manual page 118

High power dc electronic loads
Table of Contents

Advertisement

Setting Example:
Query Syntax 1:
Query Syntax 2:
Return Parameters: <aard>
Query Example:
Return Example:
PROGram:DATA:STEP
Type:
Description:
range.
Set the dwell time of sequence:
Arg9: <NRf>, 0.1ms ~ 30s, Resolution = 0.0001s, Unit =
Second.
Set the high-level of voltage specific in sequence:
Arg10: <NRf>, Refer to respective specification for valid value
range.
Set the low-level of voltage specific in sequence:
Arg11: <NRf>, Refer to respective specification for valid value
range.
Set the high-level of current specific in sequence:
Arg12: <NRf>, Refer to respective specification for valid value
range.
Set the low-level of current specific in sequence:
Arg13: <NRf>, Refer to respective specification for valid value
range.
Set the high-level of power specific in sequence:
Arg14: <NRf>, Refer to respective specification for valid value
range.
Set the low-level of power specific in sequence:
Arg15: <NRf>, Refer to respective specification for valid value
range.
Set the delay time of Pass/Fail in sequence:
Arg16: <NRf>, 0s ~ 30s, Resolution = 0.0001s, Unit = Second.
PROG:DATA:LIST 1,1,AUTO,CC,2,3.5,0.5,0.5,2,-1,-1,-1,-1,-1,-1,1
PROGram:DATA:LIST?<space><Arg1>,<Arg2>[<space><MAX |
MIN>]
Selects a program:
Arg1: <NR1>, 1 ~ 10, Resolution = 1, Unit = None.
Selects a sequence:
Arg2: <NR1>, 1 ~ N, Resolution = 1, Unit = None.
PROGram:DATA:LIST?<space><Arg1>,<Arg2>,<Arg3>,<Arg4><s
pace><MAX | MIN>
Selects a program:
Arg1: <NR1>, 1 ~ 10, Resolution = 1, Unit = None.
Selects a sequence:
Arg2: <NR1>, 1 ~ N, Resolution = 1, Unit = None.
Selects a run mode:
Arg3: <NRf>, CC(0), CR(1), CV(2), CP(3), Unit = None.
Selects the mode's range:
Arg4: <NRf>, LOW(0), MIDDLE(1), HIGH(2), Unit = None.
PROG:DATA:LIST? 2,1
PROG:DATA:LIST? 2,1 MAX
PROG:DATA:LIST? 2,1 MIN
PROG:DATA:LIST? 2,1,1,0 MAX
PROG:DATA:LIST? 2,1,1,0 MIN
2,1,AUTO,CC,HIGH,3.5,0.5,0.5,2,-1,-1,-1,-1,-1,-1,1
Channel-Specific
Set the step parameters in program. (Note: All setting parameters
in this command can't use suffix.)
Remote Operation
4-55