High Power Electronic Load 63200A Series Operation & Programming Manual
RESistance:DYNamic:T2
Type:
Description:
Setting Syntax:
Setting Parameters:<NRf+>, 10μs ~ 100s, Resolution = 10μs, Unit = Second
Setting Example:
Query Syntax:
Return Parameters:<NR2>, [Unit = Second]
Query Example:
RESistance:DYNamic:REPeat
Type:
Description:
Setting Syntax:
Setting Parameters:<NRf+>, 0 ~ 65535, Resolution = 1, Unit = None
Setting Example:
Query Syntax:
Return Parameters:<NR1>, [Unit = None]
Query Example:
RESistance:DYNamic:RISE
Type:
Description:
Setting Syntax:
Setting Parameters:Refer to respective specification for valid value range.
Setting Example:
Query Syntax:
Return Parameters:<NR2>, [Unit = A/μs]
Query Example:
RESistance:DYNamic:FALL
Type:
Description:
4-22
Channel-Specific
Set duration parameter T2 for constant resistance dynamic mode.
RESistance:DYNamic:T2<space><NRf+>[suffix]
RES:DYN:T2 10ms
RES:DYN:T2 2
RES:DYN:T2 MAX
RES:DYN:T2 MIN
RESistance:DYNamic:T2?[<space><MAX | MIN>]
RES:DYN:T2?
RES:DYN:T2? MAX
RES:DYN:T2? MIN
Channel-Specific
Set the repeat count for constant resistance dynamic mode.
RESistance:DYNamic:REPeat<space><NRf+>
RES:DYN:REP 500
RES:DYN:REP MAX
RES:DYN:REP MIN
RESistance:DYNamic:REPeat?[<space><MAX | MIN>]
RES:DYN:REP?
RES:DYN:REP? MAX
RES:DYN:REP? MIN
Channel-Specific
Set the rising slew rate of current for constant resistance dynamic
mode.
RESistance:DYNamic:RISE<space><NRf+>[suffix]
RES:DYN:RISE 2.5
RES:DYN:RISE 1A/μs
RES:DYN:RISE MAX
RES:DYN:RISE MIN
RESistance:DYNamic:RISE?[<space><MAX | MIN>]
RES:DYN:RISE?
RES:DYN:RISE? MAX
RES:DYN:RISE? MIN
Channel-Specific
Set the falling slew rate of current for constant resistance dynamic
mode.
Set the dynamic duration T2 = 10ms.
Set the dynamic duration T2 = 2s.
Set the dynamic duration T2 as
maximum value.
Set the dynamic duration T2 as
minimum value.
Set repeat count = 500
Set repeat count = maximum value.
Set repeat count = minimum value.
Set rising slew rate to 2.5A/μs.
Set rising slew rate to 1A/μs.
Set rising slew rate to the maximum
value of dynamic load.
Set rising slew rate to the minimum
value of dynamic load.