Summary Data - Agilent Technologies WireScope 350 User Manual

Hide thumbs Also See for WireScope 350:
Table of Contents

Advertisement

Chapter 5

Summary data

The WireScope 350 summary data screens present result data in tabular format.
To launch the summary data screen for any measurement, select the
measurement from the Test Data screen. Then, press the Data button.
The example above shows the summary data screen for the NEXT measurement.
In this case, the measured NEXT data did not meet the selected cabling
performance limit (Category 6 basic link) for three of the six possible pair
combinations.
The data table shows the values and margins for the worst-case frequency point
for the selected pair or pair combination. Negative margins are shown in red.
By default, the summary data screen opens with the pair or pair combination
selected which has the worst margin. You can select other pairs or pair
combinations by touching in the list box or by using the
key.
Selecting
pairs
changes the
diagram and
table to
match.
Running an Autotest 52
Diagram
show s
selected pair
combination
and w hich
end failed.
Data table
Currently
selected pair
combination
highlighted
in dark blue.