Figure 3-5: Example Pneumatic Connections Flow Diagram Through Gas Dilution - Teledyne 300E Instruction Manual

Carbon monoxide analyzer
Table of Contents

Advertisement

Model 300E Carbon Monoxide Analyzer
Instruction Manual
Some applications, such as EPA monitoring, require multipoint calibration checks
where Span gas of several different concentrations is needed. We recommend
using high-concentration, certified, calibration gas supplied to the analyzer through
a Gas Dilution Calibrator such as a T-API Model 700. This type of calibrator
precisely mixes Span Gas and Zero Air to produce any concentration level between
0 ppm and the concentration of the certified gas. This means that both the source
of Zero Air and Span Gas must be connected to the calibrator whose output is then
connected to the Pressure Span inlet on the Rear Panel of the instrument.
Figure 3-5 depicts the pneumatic set up in this sort of application of a Model 300E
CO Analyzer with Zero/Span Valve Option 50 installed.
Gas Pressure should be
regulated at
30 – 35 PSIG
Certified
Zero Air Generator
Figure 3–5: Example Pneumatic Connections Flow Diagram through
P/N 04288 REV: A
CO Gas
MODEL 701
Gas Dilution Calibrator
Source of
Source of
SAMPLE Gas
SAMPLE Gas
MODEL 700
Gas Dilution
Calibrator
VENT
VENT
VENT
Getting Started
Option 50
VENT if input is pressurized
Sample
Sample
MODEL
MODEL
Pressure Span
Pressure Span
300E
300E
Vent Span
Vent Span
IZS
IZS
Exhaust
Exhaust
31

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents