6.4.5. A1 & A2 Auto Range Mode
When the Auto Range mode is selected, the instrument reports the same data with
the same measurement span on both the A1 and A2 outputs, but automatically
switches both outputs back and forth between a user selected Low Range and a
user selected High Range depending on the level of the CO concentration. The unit
will move from Low Range to High Range when the CO concentration exceeds to
98% of the Low Range span. The unit will return from High Range back to Low
Range once the CO concentration falls below 75% of the Low Range span.
To set the ranges press following keystroke sequence:
Set the
LOW and HIGH
range value.
68
SAMPLE
RANGE = 50.0 PPM
< TST TST > CAL
SETUP C.3
CFG DAS
RNGE
PASS CLK MORE
SETUP C.3
RANGE CONTROL MENU
MODE
SET UNIT
SETUP C.3
RANGE MODE:
SNGL
SNGL DUAL
AUTO
SETUP C.3
RANGE MODE:
AUTO
SNGL DUAL AUTO
SETUP C.3
RANGE CONTROL MENU
MODE
SET
UNIT
SETUP C.3
LOW RANGE:
150.0 Conc
0
0
1
5
0
.0
SETUP C.3
HIGH RANGE:
50.0 Conc
0
0
1
5
0
.0
SETUP C.3
RANGE CONTROL MENU
MODE SET UNIT
Model 300E Carbon Monoxide Analyzer
CO =
SETUP
EXIT
EXIT
ENTR EXIT
ENTR
EXIT
EXIT
The Low and High
Ranges have separate
ENTR
EXIT
slopes and offsets for
computing the carbon
monoxide concentration.
The two ranges must be
independently calibrated.
ENTR
EXIT
EXIT Returns
to the Main
EXIT
SAMPLE Display
Instruction Manual