Getting Started
The third analog output, labeled A3 is special. It can be set by the user (see
Section 6.4.7) to output any one of the parameters accessible through the
<TST TST> keys of the units Sample Display.
The standard configuration for these outputs is mVDC. An optional Current
Loop output is available for each.
Pin-outs for the Analog Output connector at the rear panel of the instrument
are:
PIN
1
2
3
4
5
6
7
8
The default analog output voltage setting of the 300E CO Analyzer is 0 – 5
q
VDC with a range of 0 – 50 ppm.
TO change these settings, see Sections 6.4 and 6.5
q
3.
If you wish utilize the analyzer's Status Outputs to interface with a device that
accepts logic-level digital inputs, such as Programmable Logic Controllers
(PLC's) they are accessed via a 12-pin connector on the analyzer's rear panel
labeled STATUS.
24
Analog Output
A1
A2
A3
A4 (Spare)
STATUS
1
2
3
4
5
Model 300E Carbon Monoxide Analyzer
VDC Signal
V Out
Ground
V Out
Ground
V Out
Ground
V Out
Ground
+
6
7
8
D
Instruction Manual
mADC Signal
I Out +
I Out -
I Out +
I Out -
I Out +
I Out -
Not Available
Not Available
P/N 04288 REV: A