CALIBRATION PROCEDURES

7.4. Automatic Zero/Span Cal/Check (AutoCal)

The AutoCal system allows unattended periodic operation of the ZERO/SPAN valve
options by using the M300E's internal time of day clock. AutoCal operates by
executing SEQUENCES programmed by the user to initiate the various calibration
modes of the analyzer and open and close valves appropriately. It is possible to
program and run up to 3 separate sequences (SEQ1, SEQ2 and SEQ3). Each
sequence can operate in one of 3 Modes, or be disabled.
MODE NAME
Disabled
Zero
Zero-Span
Span
For each mode there are seven parameters that control operational details of the
SEQUENCE. They are:
Attribute Name
Timer Enabled
Starting Date
Starting Time
Delta Days
Delta Time
Duration
Calibrate
132
Table 7–2: AUTOCAL Modes
Disables the Sequence
Causes the Sequence to perform a Zero calibration/check
Causes the Sequence to perform a Zero and Span
concentration calibration/check
Causes the Sequence to perform a Span concentration
calibration/check
Table 7–3: AutoCal Attribute Setup Parameters
Turns on the Sequence timer
Sequence will operate after Starting Date
Time of day sequence will run
Number of days to skip between each Seq. execution
Number of hours later each "Delta Days" Seq is to be run
Number of minutes the sequence operates
Enable to do a calibration – Disable to do a cal check only
Model 300E Carbon Monoxide Analyzer
ACTION
Action
Instruction Manual