6.5.1. Voltage Output Range Selection and Offset Adjustment

To select an output signal level or add a DC offset to one of the analyzer Analog
Output Channels press:
FROM ANALOG I/O CONFIGURATION MENU
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
DIAG AIO
AIN A/C FREQUENCY: 60 HZ
SET>
EDIT
DIAG AIO
AOUT CALIBRATED: NO
< SET
SET>
CAL
DIAG AIO
CONC_OUT_2:5V, CAL
< SET SET>
EDIT
DIAG AIO
SET>
EDIT
DIAG AIO
< SET
SET>
EDIT
DIAG AIO
<SET
CAL
74
A1
1
2
Figure 6–2: Analog Output Pin Assignments
Exit to Return
ENTR
EXIT
Sample Display
EXIT
EXIT
EXIT
CONC_OUT_2 RANGE: 5V
EXIT
DIAG AIO
0.1V
CONC_OUT_2 REC OFS: 0 mV
EXIT
DIAG AIO
+
CONC_OUT_2 CALIBRATED:
NO
EXIT
DIAG AIO
<SET
Calibrates Automatically
Model 300E Carbon Monoxide Analyzer
ANALOG
A2
A3
A4
3
4
5
6
7
to the main
Press SET> to select the Analog Output
channel to be configured:
DISPLAYED AS
=
CHANNEL
CONC_OUT_1
=
A1
CONC_OUT_2
=
A2
TEST OUTPUT
=
A3
Then Press EDIT to continue
OUTPUT RANGE:
5V
1V
5V
10V CURR
ENTR
RECORD OFFSET: 0 MV
0
0
0
0
ENTR
CONC_OUT_2 CALIBRATED:
CAL
Instruction Manual
8
EXIT
These Keys set
the Signal Level
of the selected
channel
EXIT
Toggle these
Keys to set the
DC Offset
of the selected
channel
YES
EXIT
Exit to Return
to the I/O
Configuration
Menu
P/N 04288 REV: A
Pressing ENTR
records the new
setting and returns
to the previous
menu
Pressing EXIT
ignores the new
setting and returns
to the previous
menu