6.6.1. Signal I/O Diagnostic Functions

The signal I/O diagnostic mode allows access to the digital and analog I/O in the
analyzer. Some of the digital signals can be controlled through the keyboard. See
Appendix A for a complete list of the parameters available for review under this
menu.
Any Signal I/O values set while in the Signal I/O function remain as set until EXITING
the Signal I/O menu.
The Analyzer software then regains control of these signals upon exit.
To enter the signal I/O test mode, press:
=
Use the NEXT & PREV
DOWN one signal type.
Use the JUMP key go directly
(see Appendix D for Jump
Pressing the PRNT Key will send a
formatted printout over the Serial I/O Port.
RS-232/485 com ports be active and
P/N 04288 REV: A
SAMPLE
< TST TST > CAL
SETUP C.3
CFG DAS RNGE PASS CLK
SETUP C.3
COMM VARS
SETUP C.3
8
1
DIAG
keys to move UP or
PREV NEXT
DIAG I / O
to a specific Signal
PREV NEXT
address numbers.
EXAMPLE
DIAG I / O
To Jump
to Signal no 5:
0
5
SAMPLE_LED
DIAG I / O
PREV NEXT JUMP
This Requires that one of the
connected to a peripheral device.
NOTE
RANGE = 50.0 PPM
CO =
SETUP
MORE
EXIT
DIAG
HALT
EXIT
ENTER DIAG PASS: 818
8
ENTR
EXIT
SIGNAL I / O
JUMP
ENTR
EXIT
Test Signals Displayed Here
JUMP
PRNT EXIT
JUMP TO: 5
ENTR
EXIT
SAMPLE_LED = ON
ON
PRNT EXIT
Exit to Return
to the
Diag Menu
Operating Instructions
Press Exit to
return to the
SAMPLE Mode
display
See Appendix D for a
complete list of
available SIGNALS
79