The following example sets Sequence #2 to do a Zero-Span Calibration every other
day starting at 1 am on September 4, 2001, lasting 15 minutes, without calibration.
This will start ½ hour later each iteration.
Mode and
Attribute
Sequence
Mode
Timer Enable
Starting Date
Starting Time
Delta Days
Delta Time
Duration
Calibrate
The programmed STARTING_TIME must be a minimum of 5 minutes later than
the real time clock (See Section 6.3.5 for setting real time clock).
Avoid setting two or more sequences at the same time of the day. Any new
sequence which is initiated whether from a timer, the COM ports, or the contact
closure inputs will override any sequence which is in progress.
P/N 04288 REV: A
Value
2
Define Sequence #2
ZERO-SPAN
Select Zero and Span Mode
ON
Enable the timer
Sept. 4, 2001
Start after Sept 4, 2001
01:00
First Span starts at 1:00AM
2
Do Sequence #2 every other day
00:30
Do Sequence #2 ½ hr later each day
15.0
Operate Span valve for 15 min
NO
Do not calibrate at end of Sequence
NOTE
NOTE
CALIBRATION PROCEDURES
Comment
133