Teledyne 300E Instruction Manual page 130

Carbon monoxide analyzer
Table of Contents

Advertisement

CALIBRATION PROCEDURES
Analyzer enters
ZERO CAL Mode
See Table 5-1 for Z/S
Valve States during
this operating mode
This operation changes
the calculation equations
and zeroes the instrument
If you are performing a
Zero CAL CHECK DO
NOT press ENTR
Analyzer enters
SPAN CAL Mode
See Table 5-1 for Z/S
Valve States during
this operating mode
This operation changes
the calculation equations
and zeroes the instrument
If you are performing a
Span CAL CHECK DO
NOT press ENTR
If either the ZERO
or SPAN buttons
fail to appear see
Section 9 for
troubleshooting
130
SAMPLE
< TST TST > CAL
ZERO CAL M
< TST TST >
M-P CAL
< TST TST >
WARNING!
calculation equations and zeroes
ZERO CAL M
< TST TST > ZERO
SAMPLE
< TST TST > CAL CALZ
ZERO CAL M
< TST TST >
M-P CAL
< TST TST >
WARNING!
This operation changes the
calculation equations and zeroes
ZERO CAL M
tips.
< TST TST >
Model 300E Carbon Monoxide Analyzer
RANGE = 50.0 PPM
CO =??.??
CALZ
CALS
SETUP
RANGE = 50.0 PP
CO =1.23
ZERO
CONC
EXIT
RANGE = 50.0 PPM
CO =1.23
ENTR
CONC
EXIT
This operation changes the
the instrument
RANGE = 50.0 PP
CO = 0.00
CONC
EXIT
RANGE = 50.0 PPM
CO =40.58
CALS
SETUP
RANGE = 50.0 PP
CO =40.58
SPAN
CONC
EXIT
RANGE = 50.0 PPM
CO =40.00
ENTR
CONC
EXIT
the instrument
RANGE = 50.0 PP
CO =40.58
SPAN
CONC
EXIT
Instruction Manual
WAIT 10
MINUTES
Or until the
reading
stabilizes and
the ZERO button
is displayed
EXIT halts the calibration
process and returns the
unit to SAMPLE Mode.
If performing a CHECK,
press EXIT
WAIT 10
MINUTES
Or until the
reading
stabilizes and
the SPAN button
is displayed
EXIT halts the calibration
process and returns the
unit to SAMPLE Mode.
If performing a CHECK,
press EXIT
Press EXIT to Return to the
Main SAMPLE Display
P/N 04288 REV: A