6.4.6. Setting the Measurement Range Unit Type

The M300E can display concentrations in ppb, ppm, ug/m
units affects all of the COM port values, and all of the display values for all
Measurement Ranges. To change the units of Measure press:
Select the preferred units of
measure.
EXAMPLE: Switching from
PPM to UGM
Concentrations displayed in mg/m
Temperature and Pressure (STP).
Consult your local regulations for the STP used by your agency.
Once the units of Measurement have been changed the unit MUST be
recalibrated, as the "expected span values" previously in effect will no longer
be valid. Simply entering new expected span values without running the entire
calibration routine is not sufficient.
The following equations give approximate conversions between volume/volume
units and weight/volume units:
P/N 04288 REV: A
SAMPLE
< TST TST > CAL
SETUP C.3
CFG DAS
RNGE
SETUP C.3
MODE SET
UNIT
SETUP C.3
PPM PPB
UGM
SETUP C.3
PPM PPB UGM MGM
3
CO ppb x 1.25 = CO ug/m
CO ppm x 1.25 = CO mg/m
RANGE = 50.0 PPM
CO =
SETUP
PASS CLK MORE
EXIT
RANGE CONTROL MENU
EXIT
CONC UNITS: PPM
MGM
EXIT
ENTER
CONC UNITS: UGM
ENTER EXIT
NOTE
3
and ug/m
use 0°C, 760 mmHg for Standard
CAUTION
3
3
Operating Instructions
3
3
, mg/m
units. Changing
EXIT Returns
to the Main SAMPLE
Display
EXIT Returns
to the SETUP
Menu
69