Table 11-2: Test Functions - Indicated Failures - Teledyne 300E Instruction Manual

Carbon monoxide analyzer
Table of Contents

Advertisement

TROUBLESHOOTING & REPAIR PROCEDURES
Test
Functions
(
As Displayed)
Time of Day clock is too fast or slow
TIME
§ To adjust see Section 7.3.6.
§ Battery in clock chip on CPU board may be dead.
Incorrectly configured Measurement Range(s) could cause response problems with a
Datalogger or Chart Recorder attached to one of the Analog Output.
RANGE
If the Range selected is too small, the recording device will over range.
If the Range is too big, the device will show minimal or no apparent change in readings.
Indicates noise level of instrument or Co concentration of Sample Gas (see Section 11.4.2
STABIL
for causes).
If the value displayed is too high the IR Source has become brighter. Adjust the variable
gain potentiometer on the Sync/Demod Board
If the value displayed is too low or constantly changing
AND:
The CO REF is OK:
§ Failed Multiplexor on the Mother Board
CO MEAS
§ Failed Sync/Demod Board
&
§ Loose Connector or wiring on Sync/Demod Board
CO REF
The CO REF is also BAD:
§ GFC Wheel stopped or rotation is too slow
§ Failed Sync/Demod Board IR Source
§ Failed IR Source
§ Failed Relay Board
§ Failed I
§ Failed IR Photo-Detector
When the analyzer is sampling Zero Air and the ratio is too low:
§ The Reference Cell of the GFC Wheel is contaminated or leaking.
§ The alignment between the GFC Wheel and the Segment Sensor, the M/R Sensor or
both is incorrect.
§ Failed Sync/Demod Board
MR Ratio
When the analyzer is sampling Zero Air and the ratio is too high:
§ Zero Air is contaminated
§ The alignment between the GFC Wheel and the Segment Sensor, the M/R Sensor or
both is incorrect.
§ Failed Sync/Demod Board
§ Failed IR Photo-Detector
§ See Table 11-1 for SAMPLE PRES WARN
PRES
SAMPLE FL
Check for Gas Flow problems. See Section 11.3.
SAMPLE
Sample temp should be close to BENCH TEMP. Temperatures outside of the specified range
TEMP
or oscillating temperatures are cause for concern
Bench temp control improves instrument noise, stability and drift. Temperatures outside of
BENCH
the specified range or oscillating temperatures are cause for concern. See Table 10-1 for
TEMP
BENCH TEMP WARNING
Wheel temp control improves instrument noise, stability and drift. Outside of set point or
WHEEL
oscillating temperatures are cause for concern. See Table 11-1 for WHEEL TEMP
TEMP
WARNING
192

Table 11-2: Test Functions - Indicated Failures

2
C Bus
Model 300E Carbon Monoxide Analyzer
Indicated Failure(s)
Instruction Manual
P/N 04288 Rev: A