Initial Operation; Start Up - Teledyne 300E Instruction Manual

Carbon monoxide analyzer
Table of Contents

Advertisement

Getting Started

3.2. Initial Operation

If you are unfamiliar with the M300E theory of operation, we recommend that you
read Section 10 before proceeding.
For information on navigating the analyzer's software menus, see the menu trees
described in Appendix A.

3.2.1. Start Up

After electrical and pneumatic connections are made, turn on the instrument power.
The exhaust fan and pump should start. The display should immediately display a
single horizontal dash at the far, left of the display. This will last for
approximately 20 – 30 seconds as the microprocessor that is the heart of the
analyzer loads its basic operating system.
Once the CPU has completed this activity it will begin loading the analyzer firmware
and factory calibration data. During this process several progress messages,
similar to the following, will appear on the analyzer's display.
M300E CO ANALYZER
BOOT PROGRESS [XXXXXX 60% _ _ _ _ _ _ _]
The analyzer should automatically switch to SAMPLE mode after completing the
boot-up sequence and start monitoring CO gas. You should see the following
display.
SAMPLE
<TST TST>
The green SAMPLE LED on the front panel should be on, and the red FAULT LED
should be flashing, indicating a SYSTEM RESET fault. This is normal.
32
The analyzer's cover MUST be installed to ensure that the temperatures
of the GFC wheel and absorption cell assemblies are properly controlled.
SYSTEM RESET
CAL
Model 300E Carbon Monoxide Analyzer
CAUTION
CO = 00.00
CLR SETUP
Instruction Manual
P/N 04288 REV: A