Figure 10-13: Location Of Relay Board Status Led's; Table 10-1: Relay Board Status Led's - Teledyne 300E Instruction Manual

Carbon monoxide analyzer
Table of Contents

Advertisement

THEORY OF OPERATION
Status LED's
Eight LED's are located on the Analyzer's Relay board to show the current status on
the various control functions performed by the Relay Board (see Figure 6-8). They
are:
LED
Color
D1
RED
D2
YELLOW
D3
YELLOW
D4
YELLOW
D5
GREEN
D6
GREEN
D7
GREEN
D8
GREEN
174
Table 10–1: Relay Board Status LED's
Function
Watchdog Circuit
Wheel Heater
Bench Heater
Spare
Sample/Cal Gas
Valve Option
Zero/Span Gas
Valve Option
Shutoff Valve
Option
IR SOURCE
STATUS LED's
RELAY PCA
PN 04135

Figure 10-13: Location of Relay Board Status LED's

Model 300E Carbon Monoxide Analyzer
Status When Lit
Cycles On/Off Every 3 Seconds under direct control of the
analyzer's CPU.
HEATING
HEATING
N/A
Valve Open to CAL GAS
FLOW
Valve Open to SPAN GAS
FLOW
Valve Open to CAL GAS
FLOW
Source ON
DC VOLTAGE TEST
POINTS
Instruction Manual
Status When Unlit
NOT HEATING
NOT HEATING
N/A
Valve Open to SAMPLE GAS
FLOW
Valve Open to ZERO GAS
FLOW
Valve CLOSED to CAL GAS
FLOW
Source OFF
P/N 04288 Rev: A