Model 300E Carbon Monoxide Analyzer
Instruction Manual
To select the Analog Output Range Type press:
6.4.3. A1 & A2 Single Range Mode
This is the default Measurement Range mode for the analyzer.
analog outputs (A1 and A2) are set to the same range. This Measurement Range
can be set to any value between 1.0 and 1,000 ppm and is accessed through the
following keystroke sequence:
P/N 04288 REV: A
SAMPLE
RANGE = 50.0 PPM
< TST TST > CAL
SETUP C.3
RNGE
CFG DAS
PASS CLK MORE
SETUP C.3
RANGE CONTROL MENU
MODE
SET UNIT
SETUP C.3
RANGE MODE:
SNGL DUAL AUTO
Go To
Go To
Section
Section
6.4.3
6.4.4
CO =
SETUP
EXIT
EXIT
SNGL
ENTR EXIT
Go To
Section
6.4.5
Operating Instructions
EXIT Returns
to the Main
SAMPLE Display
In this mode, both
65