Teledyne 300E Instruction Manual page 127

Carbon monoxide analyzer
Table of Contents

Advertisement

Model 300E Carbon Monoxide Analyzer
Instruction Manual
STEP THREE: Perform the Zero/Span Calibration Procedure:
SAMPLE
< TST TST >
M-P CAL
< TST TST >
M-P CAL
< TST TST >
M-P CAL
< TST TST >
M-P CAL
< TST TST >
M-P CAL
< TST TST > ENTR
If the ZERO or SPAN keys are not displayed, this means that the measurement
made during that part of the procedure is too far out of the allowable range to do
allow a reliable calibration. The reason for this must be determined before the
analyzer can be calibrated. See Section 11 for troubleshooting tips.
P/N 04288 REV: A
ACTION:
Allow Zero Gas to enter
the sample port on the
rear of the instrument.
RANGE = 50.0 PPM
CO =
CAL
SETUP
RANGE = 50.0 PPM
CO =
ZERO
CONC
EXIT
RANGE = 50.0 PPM
CO =
ENTR
CONC
EXIT
ACTION:
Switch gas streams to
span gas.
RANGE = 50.0 PPM
CO =
SPAN
CONC
EXIT
RANGE = 50.0 PPM
CO =
ENTR
EXIT
SPAN CONC
RANGE = 50.0 PPM
CO =
CONC
EXIT
EXIT to Return to the Main
SAMPLE Display
CALIBRATION PROCEDURES
WAIT 10
MINUTES
Or until the
reading
stabilizes and
the ZERO button
is displayed
This operation changes the calculation
equations and zeroes the instrument
WARNING!
If you are performing a ZERO CHECK
DO NOT press ENTR
If performing a CHECK, press EXIT now.
WAIT 10
MINUTES
Or until the
reading
stabilizes and
the SPAN button
is displayed
This operation changes the calculation
equations and sets the Span of the instrument
NOTE:
In certain instances where low Span gas
concentrations are present (? 10 ppm),
both the Zero & SPAN buttons may appear
simultaneously
WARNING!
If you are performing a Span CAL CHECK DO
NOT press ENTR
If performing a CHECK, press EXIT now.
If either the ZERO or
SPAN buttons fail to
appear see Section 9 for
troubleshooting tips.
127