Model 300E Carbon Monoxide Analyzer
Instruction Manual
For example: The Low range can be set for a span of 0-150 ppm while the high
range is set for 0-50 ppm.
When the instrument's range mode is set to Dual or Auto, a second set of slope and
offset parameters is used for computing the concentration for the high range. Also,
the concentration field in the upper right hand corner of the display identifies which
concentration is currently being displayed: C1 = Low (or A1) and C2 = High (or
A2).
To set the ranges press following keystroke sequence.
P/N 04288 REV: A
SAMPLE
< TST TST > CAL
SETUP C.3
CFG DAS
SETUP C.3
MODE
SETUP C.3
SNGL
SETUP C.3
SNGL DUAL AUTO
SETUP C.3
MODE
SETUP C.3
The LOW and HIGH
0
Ranges have separate
slopes and offsets for
computing the carbon
monoxide concentration.
The two ranges must be
SETUP C.3
independently calibrated.
0
SETUP C.3
MODE SET UNIT
RANGE = 50.0 PPM
CO =
SETUP
RNGE
PASS CLK MORE
EXIT
RANGE CONTROL MENU
SET UNIT
EXIT
RANGE MODE:
SNGL
DUAL
AUTO
ENTR EXIT
RANGE MODE:
DUAL
ENTR
EXIT
RANGE CONTROL MENU
SET
UNIT
EXIT
LOW RANGE:
150.0 Conc
0
1
5
0
.0
ENTR
EXIT
HIGH RANGE:
50.0 Conc
0
1
5
0
.0
ENTR
EXIT
RANGE CONTROL MENU
EXIT
Operating Instructions
EXIT Returns
to the Main
SAMPLE Display
67