9.3.5. Flow Calibration

Once an accurate measurement has been recorded by the method described above,
adjust the analyzer's internal flow sensors by pressing:
Exit at Any
Time to Return
to main the
SETUP Menu
Toggle Keys to change
setting for
ACTUAL FLOW
until it matches the value
measured with the Flow
Meter
P/N 04288 REV: A
MAINTENANCE SCHEDULE & PROCEDURES
SAMPLE
RANGE = 50.0 PPM
< TST TST > CAL
SETUP C.3
CFG DAS RNGE PASS CLK
MORE
SETUP C.3
COMM VARS
DIAG
HALT
SETUP C.3
ENTER DIAG PASS: 818
8
1
8
DIAG
SIGNAL I / O
NEXT
Repeat Pressing NEXT until . . .
DIAG
FLOW CALIBRATION
PREV
NEXT
DIAG FCAL
ACTUAL FLOW: 751 CC/ M
0
7
5
1
EXAMPLE:
Measured Flow = 853 CC/M
DIAGFCAL
ACTUAL FLOW: 853 CC/ M
0
8
5
3
DIAG
FLOW CALIBRATION
PREV
NEXT
CO =
SETUP
EXIT
EXIT
ENTR
EXIT
ENTR
EXIT
ENTR
EXIT
ENTR EXIT
previous menu
EXIT ignores the new
setting and returns to the
ENTR EXIT
setting and returns to the
ENTR
EXIT
EXIT returns
to the
previous menu
ENTR Accepts new
previous menu
153