Table 2-1: Final Test And - Teledyne 200A Instruction Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledyne API Model 200A NO
Table 2-1: Final Test and Calibration Values
Test Values
RANGE
STABILITY
(Zero Gas)
SAMP FLW
OZONE FL
PMT
AZERO
HVPS
DCPS
RCELL TEMP
BOX TEMP
PMT TEMP
IZS TEMP
MOLY TEMP
RCEL PRES
SAMP PRES
PMT Volts
NO Conc
NO
Conc
x
PMT Volts
NO Conc
NO
Conc
x
(table continued)
Analyzer Instruction Manual, 02246, Rev. F1
X
Observed
Units
Value
PPB
PPB
cc/min
cc/min
mV
mV
V
mV
o
C
o
C
o
C
o
C
o
C
IN-Hg-A
IN-Hg-A
Electric Test & Optic Test
Electric Test
mV
PPB
PPB
Optic Test
mV
PPB
PPB
2-12
Nominal Range
50-20,000
< 0.2
500 ± 50
60 - 90
0-5000
-20 - 150
450 - 900 constant
2500 ± 200
50 ±1
8-48
7 ± 1
50 ± 0.4
315 ± 5
4 - 10 constant
25 - 30 constant
2000 ± 500
1000 ± 250
1000 ± 250
2000 ± 1000
1000 ± 500
1000 ± 500
Reference Section
5.3.4
9.1.1, Table 9-1,
9.2.5
9.3.7, Table 9-1
9.3.6, 9.3.7
9.3.8.1, Table 9-1
Table 9-1
9.3.8.5
9.3.5, 9.3.4
9.3.8.2
9.3.4.1
9.3.8.4
9.3.9
9.3.4.1
9.3.7
9.3.7
9.1.3.2
9.1.3.2
9.1.3.2
9.1.3.3
9.1.3.3
9.1.3.3