Restoring The Configuration; Setting The Laser Calibration Voltage - Kodak DirectView CR 825 Adjustments And Replacements

Hide thumbs Also See for DirectView CR 825:
Table of Contents

Advertisement

ADJUSTMENTS AND REPLACEMENTS
10DEC05
6
Remove the FLOPPY DISK after the process is complete.
AR4825-1
Page
7
Do
144 of 160
8
Do an application shutdown to return to the DESKTOP.
9
Click [OK].

Restoring the Configuration

1
Double-click the "MIMDUI".
2
Enter the password.
3
Select "Connect/Passthrough".
4
Enter the IP address of the CR 825/850 SYSTEM.
5
Select "Configure/Restore Configuration".
6
Insert the MIM BACKUP FLOPPY DISK.
7
Click [OK].
8
Remove the MIM BACKUP FLOPPY DISK.

Setting the Laser Calibration Voltage

Important
Do this procedure when installing a new OPTICAL AY.
For systems with Software Version ≥ 4.1, you must have a "Session ID" for access to
"Service Functions" and "Diagnostics". See SERVICE BULLETIN 843.
1
At the main menu, touch:
[Service Functions]
[Diagnostics]
[Laser Power]
Step 3
-
Step 6
again until all options are restored.
Additional Service Procedures